This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates analysis of camphorsulfonic acid in rat plasma on Oasis Wax.
Camphorsulfonic acid is a strong acid with a pKa of approximately 1.5. The best SPE recoveries for this type of acid are on Oasis WAX products.
Oasis WAX 10 mg 96-Well Plate |
|
---|---|
Condition: |
500 μL MeOH |
Equilibrate: |
500 μL H2O |
Load: |
500 μL (250 μL rat plasma, diluted 1:1 with 4% H3PO4) |
Wash 1: |
500 μL 2% FA, pH 2.7 |
Wash 2: |
500 μL MeOH |
Elute: |
250 μL (125 μL x 2) 5% NH4OH in MeOH |
Options: |
1. Dilute 250 μL H2O with 2% FA 2. Evaporate/ Reconstitute 3. Direct inject |
Inject: |
10 μL |
Condition: |
200 μL MeOH |
Equilibrate: |
200 μL H2O |
Load: |
100 μL (50 μL rat plasma diluted 1:1 with 4% H3PO4) |
Wash 1: |
200 μL 2% FA, pH 2.7 |
Wash 2: |
200 μL MeOH |
Elute: |
50 μL (25 μL x 2) 5% NH4OH in MeOH |
Options: |
1. Direct injection 2. Dilute with 50 μL H2O with 2% FA 3. Evaporate/ Reconstitute |
Inject: |
10 μL |
Column: |
SunFire C18 2.1 x 20 mm IS, 3.5 μm |
Mobile phase A: |
10 mM CH3COO-NH4+, pH 5.5 |
Mobile phase B: |
MeOH with 10 mM CH3COO-NH4+, pH 5.5 |
Flow rate: |
0.4 mL /min |
Injection volume: |
10 μL |
Column temp: |
Ambient |
Instrument: |
2777 Sample Manager, 1525μ Binary HPLC Pump and Quattro Premier |
ESI- source temp: |
150 °C |
Desolvation temp: |
350 °C |
Cone gas flow: |
50 L /Hr |
Desolvation gas flow: |
600 L /Hr |
Collision cell: |
2.2e-3 bar (Ar gas) |
WA60084, June 2007