This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights the analysis of methoxyverapamil using XTerra MS C18 columns.
Methoxyverapamil has been analyzed using LC-MS in this study.
            
        | 
                     Column:  | 
                     XTerra MS C18 2.1 x 30 mm, 3.5 μm (p/n: 186000398)  | 
            
| 
                     Mobile phase A:  | 
                     20mM NH4HCO3 in H2O, pH 7.0  | 
            
| 
                     Mobile phase B:  | 
                     ACN  | 
            
| 
                     Isocratic mobile phase composition:  | 
                     57%A; 43%B  | 
            
| 
                     Flow rate:  | 
                     0.2 mL/min to MS  | 
            
| 
                     Injection volume:  | 
                     20 μL of 100 pg/μL  | 
            
| 
                     Temperature:  | 
                     Ambient  | 
            
| 
                     Detection:  | 
                     MS ESI+, SIR 485.4  | 
            
| 
                     Instrument:  | 
                     Alliance 2795 HT, Micromass ZQ  | 
            
| 
                     MS sytem:  | 
                     Micromass ZQ  | 
            
| 
                     Ion source:  | 
                     ESI+  | 
            
| 
                     Capillary:  | 
                     3.0 kV  | 
            
| 
                     Cone:  | 
                     45 V  | 
            
| 
                     Extractor:  | 
                     3.0  | 
            
| 
                     Source temp.:  | 
                     150 °C  | 
            
| 
                     Desolvation temp.:  | 
                     350 °C  | 
            
| 
                     Cone gas flow:  | 
                     60 L/hr  | 
            
| 
                     Desolvation gas flow:  | 
                     500 L/hr  | 
            
| 
                     LM Resolution:  | 
                     15  | 
            
| 
                     HM Resolution:  | 
                     15  | 
            
| 
                     Ion energy:  | 
                     1  | 
            
| 
                     Multiplier (V):  | 
                     650  | 
            
            
        WA20738.068, June 2002