This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates analysis of amitrole in river water.
The compound analyzed in this study is amitrole.
            
        | 
                     Column:  | 
                     Waters Spherisorb Silica, 2 x 100 mm, 3 µm  | 
            
| 
                     Part number:  | 
                     PSS832022  | 
            
| 
                     Mobile phase A:  | 
                     Acetonitrile  | 
            
| 
                     Mobile phase B:  | 
                     0.05% phosphoric acid  | 
            
| 
                     Flow rate:  | 
                     200 µL/min  | 
            
| 
                     Instrument:  | 
                     Alliance 2695 Separations Module  | 
            
            
        | 
                     MS instrument:  | 
                     Micromass ZQ  | 
            
| 
                     Ion source:  | 
                     ESI+  | 
            
| 
                     Mode:  | 
                     Multiple selected-ion recording  | 
            
| 
                     Source temp.:  | 
                     150 °C  | 
            
| 
                     Desolvation temp.:  | 
                     450 °C  | 
            
| 
                     Drying gas:  | 
                     500 L/hr  | 
            
| 
                     Cone:  | 
                     50 V  | 
            
            
        
            
        1 Calculated against standard prepared in matrix. Matrix suppres sion was approximately 10%
WA31764.41, June 2003