This study evaluates the sensitivity and mass measurement of the Waters Micromass LCT Premier Mass Spectrometer when configured with the IonSABRE APCI probe. Using positive ion mode APCI, excellent mass accuracy is achieved routinely with continuum and real-time centroid data acquisition. The system used is comprised of a Waters Alliance HT 2795 Separations Module, Symmetry C18 Column and LCT Premier Mass Spectrometer.
Although the vast majority of LC-MS applications are per formed with electrospray ionization, atmospheric pressure chemical ionization (APCI) is often chosen when compounds under analysis require extension of the polarity range (typically more nonpolar).
In this study, APCI is coupled with the newly developed LCT Premier orthogonal acceleration time of flight (oa-TOF) mass spectrometer (Figure 1). This benchtop mass spectrometer incorporates new hardware and software control technology to meet the analytical demands of pharmaceutical, environmental and clinical applications.
This study uses a set of standard compounds to evaluate LCT Premier performance with APCI using both continuum and real-time exact mass centroid data acquisition. The high duty cycle of the TOF is utilized for qualitative studies, generating full spectra at high mass accuracy (<3 ppm RMS). This mass measurement accuracy provides an extra degree of information that aids in the interpretation of data.
HPLC system: |
Waters Alliance HT 2795 Separations Module |
Column: |
Waters Symmetry C18, 4.6 mm x 100 mm, 3.5 μm |
Flow rate: |
1 mL/min |
Mobile phase: |
A: H2O, B: MeCN |
Isocratic: |
70% A: 30% B |
Mass spectrometer: |
Waters Micromass LCT Premier oa-Tof |
Ionization mode: |
APCI Corona +ve = 10 μA |
Sample cone voltage: |
100V |
Reference mass: |
Propranolol , [M+H]+ = 260.1650, post-column tee-in |
Acquisition parameters: |
100 - 1000 m/z 1 spectrum/second 0.1 second inter scan delay 5500 FWHM (V-mode) 10000 FWHM (W-mode) |
Standards were prepared in 50/50 acetonitrile/water. The compounds used in this study are presented with the elemental formula in Table 1.
Presented in Figure 2 is an example of a typical APCI calibration using sodium formate acquired using the LCT Premier auto-calibration procedure.
Table 1 lists the standard compounds used to test the LCT Premier APCI performance. The standards were chosen in order to illustrate the mass resolution and accuracy that can be obtained for small molecule applications. The data acquisition mass range utilized was m/z 100-1000. The standards used range from m/z 152 to m/z 609. The resolution achieved for progesterone (m/z 331) at >5000 FWHM (V-mode) and >10000 FWHM (W-mode) is presented in Figure 3.
To deter mine the sensitivity of the LCT Premier in positive APCI mode, consecutive 10 μL loop injections of 17α hydroxyprogesterone (1ng/μL) were made and the signal to noise determined. Over six injections, an average S/N= 452:1 was achieved, with a coefficient of variation of 3.6% as shown in Figure 4. The sensitivity in negative APCI mode was determined with consecutive 10 μL loop injections of sulphadimethoxine (1ng/μL).The signal to noise determined over six injections gave an average S/N= 2992:1, with a coefficient of variation of 3.6% as shown in Figure 5.
Using loop injections of a mixture of five standards, six consecutive injections were performed to determine the reproducibility of the response and the exact mass measurement achieved. Initially, continuum data was acquired and centered. This is illustrated in Figure 6. The five masses of interest are labelled 1 to 5. It can be seen in Figure 7 that excellent exact mass measurement has been obtained. The calculated masses are shown alongside the mass measured. The error varied from –2.58 ppm to 2.11 ppm, giving an RMS error of just 1.61 ppm over the mass range acquired.
The RMS ppm errors for six consecutive injections of each compound are shown in Table 2. The minimum RMS ranged from 1.05 ppm for m/z 152 to 3.24 ppm for m/z 331.
Mass Measurement Accuracy in Centroid Mode The system was tested further by repeating the experiment in centroid mode. Presented in Figure 8 is an example of the exact mass measurement obtained for APCI centroid mode. The five standards used are labelled 1 to 5. Standard number 3 (17α hydroxyprogesterone) was selected and entered into the elemental composition calculator to generate the most probable elemental composition and the mass measurement error. As can be seen from Figure 9, only one probable elemental composition was produced, with a mass measurement error of only 1.1 ppm.
It can be seen in Figure 10 that excellent exact mass measurement has been obtained for all five standards. The calculated accurate masses are shown alongside the mass measured. The error varied from 0.8 ppm to 1.87 ppm, giving an RMS error of only 1.25 ppm over the mass range acquired. The RMS ppm errors for six consecutive injections of each compound are shown in Table 3. The minimum RMS ranged from 3.7 ppm for m/z 152 to 0.9 ppm for m/z 609.
Dynamic Range Shown in Figure 11 is an illustration of the linear response obtained for sulphadimethoxine using DRE (Dynamic Range Enhancement). Four orders of magnitude dynamic range are shown for sulphadimethoxine, where concentrations ranging from 1 pg/μL to 10000 pg/μL were injected on column. A correlation coefficient of 0.9997 was achieved.