Determination of Aromatic Hydrocarbon Types in Middle Distillates with the Alliance HPLC System and RI Detection According to ASTM D6591 (IP548)

Library Number:
APNT135078152
Part Number:
720007073EN
Author(s):
Rachel Sanig, Michael Jones
Source:
Waters Corporation
Content Type:
Application Notes
Content Subtype:
Application Notes
Related Products:
 
 
 

The Waters Alliance HPLC System, with the Waters Two-Position Six-Port Switching Valve, coupled to the Waters 2414 Refractive Index Detector, is shown to greatly exceed the performance requirements for the application of the ASTM D6591 (IP548) standard method for the determination of aromatic hydrocarbons in middle distillates. The instrument set up demonstrates compliance with the ASTM standard method, meeting the linearity, resolution, and precision criteria, with calibration coefficients that demonstrate an excellent linear response for quantification of aromatic hydrocarbons.


Title Format File Size
Download PDF PDF 421.06kB