For research use only. Not for use in diagnostic procedures.
This is an Application Brief and does not contain a detailed Experimental section.
This Application brief demonstrates the simultaneous and high sensitivity analysis of plasma metanephrines by LC-MS for clinical research.
Measurement of several important plasma biogenic amines
Development of LC-MS methods for simultaneous measurement of plasma metanephrines has been challenging. Efficient sample pre-treatment strategies have yet to be defined and measurement of these molecules can be complicated by the difficult chromatographic separation of metanephrines such as 3-methoxytyramine (3-MT), from other metanephrines. Development of a good separation has proven problematic and can lead to overestimation of metanephrine levels. Many laboratories also do not measure 3-MT as the assays they use may not have the analytical sensitivity to measure this low-level analyte. To effectively study these important biogenic amines, a research method to measure metanephrine, normetanephrine and 3-MT simultaneously is needed.
 
            
        | LC System: | ACQUITY UPLC with ACQUITY UPLC Online SPE Manager | 
| Mass Spectrometer: | Xevo TQ-S | 
| Column: | Atlantis HILIC, 2.1 x 50 mm, 3 μm | 
| Sample Preparation: | ACQUITY UPLC Online SPE Manager (OSM) | 
| SPE: | MassTrak WCX OSM Cartridge | 
Plasma samples were diluted 1:1 with deuterated internal standard and then centrifuged through a 10 k MW cutoff centrifugation spin filter device to remove proteins from the sample. 
After filtration, an aliquot of sample was injected into the online SPE equipped LC-MS system. SPE was performed by the online SPE system as follows:
 
            
        After SPE, samples were analyzed by LC-MS using the following gradient conditions:
 
            
         
            
         
            
         
            
         
            
         
            
        In this study, a clinical research method has been developed for the simultaneous measurement of three plasma biogenic amines. The method utilizes LC-MS and an online SPE system. This combination of SPE sample preparation coupled with the analytical power of LC-MS is able to deliver an efficient assay for these important plasma metanephrines.
The method developed here provides:
720004613, September 2013