This is an Application Brief and does not contain a detailed Experimental section.
| Column: | ACQUITY UPLC BEH HILIC, 2.1 x 50 mm, 1.7 µm | 
| Part Number: | 186003460 | 
| Mobile Phase A: | 10 mM NH4COOH, 0.2% HCOOH in 50:50 ACN: H2O | 
| Mobile Phase B: | 10 mM NH4COOH, 0.2% HCOOH in 90:10 ACN: H2O | 
| Flow Rate: | 0.529 mL/min | 
| Injection Volume: | 5.0 µL | 
| Sample Concentration: | 5 µg/mL | 
| Sample Diluent: | 73:25:2 ACN:MeOH:H2O with 0.2% HCOOH and 5 µM HCl | 
| Temperature: | 30 C | 
| Instrument: | Waters ACQUITY UPLC with SQ Mass Detector | 
| Time (min) | %A | %B | 
|---|---|---|
| 0.0 | 0.1 | 99.9 | 
| 1.65 | 0.1 | 99.9 | 
| 4.49 | 99.9 | 0.1 | 
| 4.54 | 0.1 | 99.9 | 
| 5.05 | 0.1 | 99.9 | 
| Ionization Mode: | ES+ | 
| Capillary: | 3.0 kV | 
| Cone: | 20 V | 
| Extractor: | 3 V | 
| RF Lens: | 0.3 V | 
| Source Temperature: | 150 °C | 
| Desolvation Temperature: | 350 °C | 
| Cone Gas Flow: | 50 L/Hr | 
| Desolvation Gas Flow: | 700 L/Hr | 
| SIR: | 136.1 m/z Homocysteine 150.2 m/z Methionine | 
 
            
         
            
         
            
         
            
        WA60134, August 2009