In this application note, we demonstrate the process of metabolite identification with a simultaneous determination of metabolic stability using MetaboLynx.
In modern drug discovery, data obtained from in vitro metabolism studies are used in a prospective manner to choose lead drug candidates. A promising drug candidate should possess commercially acceptable pharmacokinetic (PK) properties. Early metabolic information is useful to guide the decision-making process for structural modifications to achieve the desired PK properties. In addition, the ability to evaluate in vitro metabolic stability in human liver microsomes is useful for predicting the clearance and half-life in humans of new chemical entities.
For discovery metabolism studies, throughput remains one of the most critical factors to consider during method development. Metabolism data should be generated and interpreted rapidly; fast turnaround times allow chemists to incorporate the data into synthesis of the next compound series.
Sample throughput in drug metabolism can be increased in two stages:
Data processing and interpretation has always been a major bottleneck for drug metabolism studies. MetaboLynx is a software tool that automates the process of peak detection and data interpretation. Results are presented in a browser with an easy-to-review format.
In addition, if the incubation was performed with multiple time points, MetaboLynx also automatically performs the metabolic stability calculation while performing the automated peak detection and data interpretation. This metabolic stability information is embedded within the MetaboLynx browser for easy review.
In this application note, we demonstrate the process of metabolite identification with a simultaneous determination of metabolic stability using MetaboLynx. The example used here is an in vitro microsome incubation of buspirone. All sample analyses were performed by using the ACQUITY UPLC System with the Quattro Premier XE Mass Spectrometer (Figure 1).
The parent drug buspirone was incubated with human and rat liver microsomes at a 100 μM level. The incubation was carried out at 37 °C, in a solution of 50 mM potassium phosphate adjusted to pH 7.4 containing the appropriate cofactors. The reaction was terminated with two volumes of cold acetonitrile to one volume of sample. The samples were stored frozen at -20 °C and diluted 1:2 ratio prior to UPLC-MS analysis.
LC System: |
Waters ACQUITY UPLC System |
Column: |
ACQUITY UPLC BEH C18 1.7 μm, 2.1 x 150 mm |
Column temp.: |
90 °C |
Mobile phase A: |
Water + 0.1% Formic acid |
Mobile phase B: |
Acetonitrile + 0.1% Formic acid |
Time(min) |
Flow(mL/min) |
%A |
Curve |
---|---|---|---|
0.00 |
0.800 |
95.0 |
|
5.25 |
0.800 |
30.0 |
6 |
5.70 |
0.800 |
0.0 |
1 |
8.00 |
0.800 |
95.0 |
1 |
Mass spectrometer: |
Waters Quattro Premier XE |
Ionization mode: |
Electrospray positive |
Capillary voltage: |
3 kV |
Cone voltage: |
40V |
Source temp.: |
130 °C |
Desolvation temp.: |
470 °C |
Acquisition mode: |
MS full scan |
A UPLC-MS full-scan experiment was performed for the incubated samples to obtain an initial screening of the metabolites. Figure 2 shows the extracted ion chromatograms of buspirone (m/z 386), hydroxybuspirone (m/z 402), and dihydroxybuspirone (m/z 418).
Previous reports1, 2 have indicated that buspirone metabolites include six hydroxybuspirone and as many as eight dihydroxybuspirone. Results displayed in Figure 2 showed six hydroxybuspirone plus at least 13 dihydroxybuspirone. Further experiments are required to positively confirm the dihydroxybuspirones (beyond scope of this application note).
A complete report of the UPLC-MS full-scan experiments can be reviewed in the MetaboLynx browser, as shown in Figure 3.
In Figure 3, the upper-left corner displays the plate map with the location of samples injected. In this example, vial 1 (dark green) was the control sample, and vials 2 to 5 contain the incubated samples at incubation time points of 15 min, 30 min, 60 min, and 90 min.
The top panel displays the list of identified metabolites. The bottom-right panel displays chromatograms. Of the two chromatograms shown here, the upper chromatogram is the combined trace of all identified metabolites plus the parent drug. Each peak is labeled with the metabolite’s name, formula weight, and retention time. The lower chromatogram is the selected ion chromatogram of the hydroxybuspirone at m/z 402. The bottom-left panel shows the spectrum that corresponds to one of the hydroxylated metabolites of buspirone. This specific example shows the spectrum of the hydroxybuspirone with retention time of 1.7 min.
By clicking on the metabolic stability button (circled in red in Figure 3), a separate window is opened to display the metabolic stability results (Figure 4).
The top panel in Figure 4 lists all of the identified metabolites, including the peak area of each metabolite at each time point. The half-life of the parent drug was calculated and shown on the top of the panel. In this example, the half-life for buspirone was 56 min. The bottom panel shows the plot of the disappearance for the parent drug as well as the appearance for metabolites.
We have demonstrated an entire information workflow that uses MetaboLynx to automatically perform data processing, result interpretation, as well as report generation. By performing a single UPLC-MS injection for each sample, we were able to obtain multiple levels of information. This includes the identification of potential metabolites, calculation of the metabolic stability such as half-life of the parent drug, as well as the rate of appearance of the metabolites.
As a result, the significant time-savings reaped by both the fast and complete UPLC-MS analysis as well as the resulting streamlined data analysis enables scientists to be more adept as they make decisions about the metabolic stability of subsequent compound synthesis in these high-throughput screening drug discovery activities.
720002762, August 2008.