Chemistry |
C18 |
Separation Mode |
Reversed Phase |
Particle Substrate |
Silica |
pH Range Min |
2 pH |
pH Range Max |
8 pH |
Temperature Limits |
45 C |
Maximum Pressure |
18000 psi (1240 Bar) |
Endcapped |
Yes |
Bonding Technology |
T3 |
Silanol Activity |
Medium |
Particle Shape |
Spherical |
Particle Size |
2.5 µm |
Endfitting Type |
Parker-style |
Pore Size |
100 Å |
Format |
Column |
Surface Area |
230 |
System |
UHPLC, UPLC |
Particle Technology |
HSS |
USP Classification |
L1 |
Inner Diameter |
3 mm |
Length |
75 mm |
Carbon Load |
11 % |
eCord |
Yes |
UNSPSC |
41115709 |
Brand |
XSelect |
Product Type |
Columns |
Units per Package |
3 pk |
XSelect HSS T3 Column, 100Å, 2.5 µm, 3 mm X 75 mm, 3/pk
The Waters XSelect HSS T3 Column, which is a low-coverage C18 phase compatible with a 100 percent aqueous mobile phase, is a clear first choice when developing separations for polar and non-polar substances. While being completely scalable and adaptable to UPLC separations, the lab equipment demonstrates ultra-low MS bleed.
Add the Waters XSelect HSS T3 Column to your lab for improved selectivity for acidic, alkaline, and neutral chemicals. You can take advantage of the XSelect column's unique mix of efficiency and selectivity to improve the resolution of very complex samples, saving you time and effort when constructing a chromatographic test. When compared to BEH and CSH particles, the HSS particle has a greater silanophilicity, which provides benefits such as improved polar component retention and dramatically different selectivity. The XSelect HSS T3 Column also offers excellent reproducibility and transferability.
The XSelect HSS T3 Column, like all Waters lab equipment, is made to the highest industry standards in a cGMP, ISO 9001 certified plant using ultra-pure reagents. Additionally, each column is tested individually to ensure that you only receive high-quality items with consistent results.
The XSelect HSS T3 Column has higher retention for polar chemicals than the standard C18 phase. When using up to 100% aqueous mobile phases, the reversed-phase analytical column outperforms the competition. The XSelect HSS T3 Column provides the most retention capability when employing typical reversed-phase mobile phases for combinations of polar and non-polar chemicals.
We recommend that you also use the XSelect HSS T3 XP VanGuard Cartridge, 100Å, 2.5 µm, 2.1 mm X 5 mm, 3/pk in conjunction with the analytical column. The cartridge is manufactured using the exact same packing materials that are used in the column, making them ideally suited. It functions by preventing any physical or chemical contaminants from entering the mobile phase, making the results more dependable. It also helps to enhance the lifetime of the XSelect Column.
What Is The pH Range Offered By The Waters XSelect HSS T3 Column?
The XSelect HSS T3 Column listed here offers an operating pH range of 2-8.
Are There Any Temperature Limits For The Waters XSelect HSS T3 Column?
This particular Waters XSelect HSS T3 Column is able to function with up to 45 C.