This application demonstrated the automated and fast method development capability of the ACQUITY UPLC System with 2D Technology for the analysis of illicit drugs in urine.
The quantification limit for all illicit drugs was set at 1 ppb using a 1 mL urine sample. The micro extraction protocol offered the option to evaluate several elution parameters in a short time period. The elution optimization was completed within a 4 hrs hands-on work and the 2D LC results were analyzed using an over-night run multi-methods sample list (18 hrs). With the extraction protocol optimized, the final protocol produced a clean extract in 15 minutes without any evaporation to dryness and reconstitution into initial mobile phase conditions. The reversed-phase/ion-exchange extraction protocol gave a 90% recovery average for the 12 illicit drugs in urine.
The field of forensic toxicology encompasses several disciplines, such as analytical chemistry, and pharmacology with the main objective of assisting legal investigation of death, poisoning and drug use. The use of spot testing is perhaps the most known technique in forensic analysis for rapid results. For a complete legal identification, a more robust methodology is required and the current trend in forensic laboratories is the use of liquid chromatography coupled with mass spectrometry (LC-MS or LC-MS/MS). However, to achieve satisfactory results most often with a trace level detection requirements, extensive and time consuming sample preparation protocols are required to reach sub ppb levels. In recent years, advances in analytical capabilities with hyphenated instrumentation platforms have enabled analytical sensitivity and efficiency to detect trace level analytes. However, the bottleneck resides with the sample preparation techniques which have not yet been modernized. Although these techniques are “tried and true”, there is much room for improvement. Utilizing Multi-Dimensional chromatography as a micro extraction technique can decrease sample preparation time while enhancing the separation integrity observed with current single dimension chromatography techniques.
The extraction process was performed using a reversed-phase solid-phase extraction in 1D, 2D optimized, 2D sequential and cumulative elution modes. The concept of sequential 2D extraction was designed to capture the retention behaviour of a target analyte in response to various extraction parameters (sorbent strength, elution polarity, solubility... etc.). Therefore, optimized conditions can be selected to excise a region of interest during extraction. Twelve illicit drugs were spiked in human urine and extracted using three extraction protocols for performance evaluation. The chosen 2D chromatography conditions used in this application were identified using a 6 x 6 automated methods development protocol (144 methods total). The manual extraction of urine samples was completed in less than 20 minutes. The analysis was performed using 200 μL of the final organic solvent (MeOH) extracts. The LOD for all drugs was measured at 100 ppt from a 1 mL sample volume. Several analytes showed excellent signal at 10 ppt.
Two MRM transitions (quantification and confirmation) for all illicit drugs were selected and optimized. The MRM conditions are listed in Table 1. For this application, finding the optimum extraction and chromatographic condition for this multi-residue analysis poses a difficult challenge.
The chromatographic conditions were tested on several trapping chemistries (Oasis HLB, XBridge C18, and XBridge C8) and separation chemistries (ACQUITY UPLC BEH C18 and HSS T3). The loading (low pH, high pH and neutral pH) and eluting mobile phase (MeOH + 0.5% formic acid and ACN + 0.5% formic acid) were also optimized using an automated process.
The extraction process was performed using a reversed-phase/ion-exchange sorbent with a 3 cc Oasis MCX SPE barrel. The sorbent was conditioned by using 5 mL of methanol followed by 5 mL of water. The urine samples (1 mL) were extracted with 1 mL acetonitrile to precipitate residual protein material. The mix was then centrifuged at 4000 rpm for 5 min. The supernatant was diluted with 48 mL of Millipore water and the entire volume was loaded at a flow rate of 10 mL/min. The cartridge was washed with 5 mL water with 2% formic acid, followed by 5 mL of methanol. The illicit drugs were eluted with 1 mL of 100% methanol with 2% ammonium hydroxide.
Column: |
Oasis HLB 20 μm |
Loading: |
MilliQ water (pH 7, no additives) |
Flow rate: |
2 mL/min |
At-column dilution: |
5% (0.1 mL/min pump A and 2.0 mL/min pump B) |
UPLC system: |
ACQUITY UPLC 2D configured for “Trap and Elute” with At-Column Dilution |
Runtime: |
10 min |
Column: |
ACQUITY UPLC BEH C18, 2.1 x 50 mm, 1.7 μm |
Column temp.: |
60 °C |
Mobile phase A: |
Water + 0.5% formic acid |
Mobile phase B: |
Acetonitrile + 0.5% formic acid |
Elution: |
5 minute linear gradient from 5% (B) to 95% (B) |
Flow rate: |
0.500 mL/min (pump C) |
Injection volume: |
200 μL |
MS system: |
Xevo TQD |
Ionization mode: |
ESI positive |
Capillary voltage: |
3.0 kV |
Cone voltage: |
30.0 V |
Source temp.: |
150 °C |
Desolvation temp.: |
550 °C |
Desolvation gas: |
1100 L/hr |
Cone gas: |
50 L/hr |
This application demonstrated the disruptive nature of the ACQUITY UPLC with 2D Technology with the Xevo TQD Mass Spectrometer for the analysis of illicit drugs in urine. The at-column dilution 2D configuration is a three pump variant using two quaternary and one binary pump. The chromatography performance utilizing at-column dilution is shown in Figure 1 for two illicit drugs (cocaine and amphetamine) between aqueous and organic extracts. The bottom chromatograms are aqueous extracts and clearly show lower signal when compared to the methanol and acetonitrile extracts. At trace level concentration (<ppb), a target analyte dissolved in an aqueous matrix will interact by ion-exchange with silanols in glass vials. When the same analyte is dissolved in an organic solvent (methanol, acetonitrile or acetone), ion-exchange is disrupted, resulting in a drastic signal increase.
The trap and elute 2D configuration is usually constructed with two circuits. The first one is to load the analyte of interest onto a trap material and the second circuit is used for backflush elution onto a high resolution analytical column. This configuration is limited to aqueous samples only. In recent development, adding a third stream (3 pump variant) has several advantages over the two stream design.1,2 The optimization process for multi-dimensional chromatography instruments is a crucial process and can be time consuming to find optimum elution conditions. With 2D platforms, there are four key parameters to optimize that include (loading pH, trapping strength, elution strength and separation selectivity. Those parameters are split between the trap and the analytical column. For efficient trapping conditions, the loading pH, flow rate and retention chemistry are essential for ensuring a tight and narrow injection band on the trap sorbent. After the loading phase is completed, the target analytes are back flush eluted onto an analytical column for high resolution separation. For optimum separation, the usual chromatography parameters (flow rate, solvent, pH, and chemistry) are also optimized for peak performance. As such, finding the best chromatographic conditions can be quite difficult. By selecting the most common elution (Si-C4, Si-C8, Si-C18, polymer, ion-exchange... etc.) , polarity (MeOH, ACN, acetone... etc.), dilution ratio (low vs. high) and pH (low, neutral and high) well over 400 permutations can be selected. A rational approach to select the most common parameters can be autonomously screened during an overnight run (18 hours). The At-column dilution ratio is calculated from the flow rate of the loader and dilutor pump. This ratio is crucial and if not set properly can lead to peak distortion. In most instances a dilution ratio set at 5% is an excellent starting point for most analytes in methanol, acetonitrile or acetone extract (decreasing polarity value). With a dilution ratio set at 5%, the analyte (cocaine) produce a Gaussian peak shape when dissolved in methanol and acetonitrile. However, with a less polar solvent like acetone, the analyte shows signs of peak distortion (see Figure 2). The distortion can be corrected by decreasing the dilution ratio. When set at 2.5%, all three extracts show excellent peak shape. With a parallel 2D configuration, the loading phase is done on the first trap, while the second trap loaded from the previous injection is eluted toward the analytical column.
Once the dilution ratio is selected, the next phase optimization of the chromatography (trapping and separation). As mentioned earlier, the task can be difficult to manage because of the huge amount of permutations. As a starting point, the 3 x 3 grid is designed to evaluate the effect of pH loading vs the trap retention strength while keeping the elution condition at a fixed value. By selecting three pH values (3, 7, and 10) and trap chemistries (C8, C18 and HLB), a 3 x 3 grid will yield 9 permutations. With a 10 minutes run time, the results are tabulated in 4.5 hours (See Figure 3). The pH and retention strength for the trap are the primary parameters for efficient retention. By adding a low and high pH elution, a 3 x 6 grid gives 18 permutations (9 hours). The elution polarity is also a key elution parameter which increases the grid to a 6 x 6 with 36 permutations and can be completed in 18 hours. For representation purpose, a color code was selected to create a database between methods and end results. As seen in Figures 4 and 5, the end results for the 36 permutations are tabulated for each illicit drugs.
The yellow and red results are also crucial data points as they can give a clear insight as to the target analyte retention behavior during the extraction process. With the at-column dilution option, aqueous and organic extracts can be utilized to give additional information regarding an analyte’s solubility. Overall, as more conditions are optimized, the elution behavior of an analyte becomes clearer. As a result, the illicit drugs were split into two methods (see Figure 6 and 7).
After the optimization process, the work will concentrate on the extraction process. This application targets illicit drugs in urine. With At-column dilution enabled, the extraction optimization can be streamlined by using a micro extraction protocol. As a starting point, the workflow will start with the reversed-phase protocol listed in Figure 8. The initial sample volume is limited at 1 mL using a 1 cc cartridge with a 1 mL elution volume. The elution is transferred directly (no evaporation/reconstitution) into a 2 mL vial. In this application, since the micro extraction can be completed in less than 15 minutes, the workflow can evaluate several elution profiles in a short time. In this case, the micro extraction protocol evaluated the effect of solvent polarity and pH on the recovery for the illicit drugs. The results for amphetamine with various elution conditions are depicted in Figure 9. The results were compared to an un-extracted standard and a recovery percentage was calculated for all illicit drugs (see Figure 10).
From the recovery results, two elution conditions were selected for first round of extraction with urine sample. In this instance, a reversed-phase sorbent was used for the extraction and eluted with 1 mL of methanol (first set) and 1 mL acetonitrile (second set). From the aqueous screening results, a urine sample was tested with two elution conditions (see Figure 11). The acetonitrile added to the urine sample was use to precipitate all remaining protein material. As seen in Figure 11, a 1:1 ratio urine/acetonitrile created a clear supernatant. After centrifugation, the supernatant is simply diluted to 50 mL, thus reducing the percentage of acetonitrile to 1%. At that organic percentage, the illicit drugs are not at risk of breakthrough. In radar mode, a full scan data and multiple reaction monitoring channels (MRM’s) can be acquired during the same run. This feature gives a unique insight to the quality of the extraction procedure and showcases the amount of potential interferences (scan mode) versus a target analyte signal to noise ratio in MRM mode.
As to be expected with a generic SPE method, a urine sample shows a higher level of interferences that an LC-MS/MS grade water (used as a reference for this application) (Figure 12). The recovery results (Figure 13) demonstrate an interference closely eluting to the target analyte. Overall, the 3 target analytes were un-detected using a reversed-phase extraction protocol when using methanol as well as acetonitrile (see Table 2). The recovery values with acetonitrile for 5 target analytes are below the acceptable range of 70% – 120%.
In order to reach better performance, the extraction protocol efficiency will need to be upgraded with a dual reversed-phase/ion-exchange protocol (see Figure 14). In this instance, since the elution with acetonitrile gave better results than methanol, the RP/IE protocol was evaluated with acetonitrile only. The stronger washes of the RP/IE protocol removed more interferences than the RP 1D and RP 2D protocols (Figure 15). The chromatograms in Figure 16 shows the trace signal for amphetamine using the RP 1D, RP 2D and RP/IE, respectively. The results in Table 3 (Getting this from the author) shows the RP/IE protocol producing excellent results for the entire mix of illicit drugs and also capturing un-detected analytes with the RP 1D protocol.
This application demonstrated the automated and fast method development capability of the ACQUITY UPLC System with 2D Technology for the analysis of illicit drugs in urine. The quantification limit for all illicit drugs was set at 1 ppb using a 1 mL urine sample. The micro extraction protocol offered the option to evaluate several elution parameters in a short time period. The elution optimization was completed within a 4 hrs hands-on work and the 2D LC results were analyzed using an over-night run multi-methods sample list (18 hrs). With the extraction protocol optimized, the final protocol produced a clean extract in 15 minutes without any evaporation to dryness and reconstitution into initial mobile phase conditions. The reversed-phase/ion-exchange extraction protocol gave a 90% recovery average for the 12 illicit drugs in urine.
720005533, February 2017