This application note details the analysis of pesticide residues and pharmaceuticals of concern in drinking water samples present at trace levels (sub ppb).
Waters UPLC with On-line SPE Technology provides a number of benefits compared to traditional off-line sample prep:
Target contaminants in water are usually present at trace levels (sub ppb). In order to bring the final concentration of analytes into the detectable range for a chosen detector (e.g. UV, ELSD, MS), a pre-concentration stage is required. An enrichment factor is calculated from the original sample volume and the final extract volume. For example, an initial sample volume of 500 mL is concentrated to a final volume of 1 mL for a 500:1 enrichment. It is common to have methods starting with higher volumes, such as 1 L, and a final extract as low as 500 μl. These methods require extensive manual labor with loading, evaporation, and reconstitution steps that may take several hours.
This application note details the analysis of pesticide residues and pharmaceuticals of concern in drinking water samples present at trace levels (sub ppb). Sample volumes of less than 20 mL were used to achieve the same trace level results as 1 L samples using the off-line approach. Samples can be collected in the field directly into 20 mL vials.
The UPLC with On-line SPE Technology used for these analyses provides a number of advantages:
Column: |
Oasis HLB Direct Connect HP 20 μm, 2.1 x 30mm Column |
Load/wash: |
2% NH4OH (both for pesticides and pharmaceuticals) |
Recondition: |
2% NH4OH in methanol |
LC system: |
Waters ACQUITY UPLC System |
Column: |
ACQUITY UPLC BEH C18 2.1 x 50 μm, 1.7 μm |
Mobile phases for pharmaceutical method: |
A: 20 mM NH4HCO2, pH 3.2 B: 20 mM NH4HCO2 in 50:50 methanol/acetone, pH 3.2 |
Mobile phases for pesticide method: |
A: 0.5% HCO2H B: 0.5% HCO2H in 80:20 methanol/acetonitril |
Gradient: |
5% B to 95% B in 5 min; 95% B held for 3 min before returning to initial conditions |
Total run time: |
13 min |
MS system: |
ACQUITY TQD |
Capillary voltage: |
3.5 kV |
Source temp: |
140 °C |
Desolvation temp: |
350 °C |
Desolvation gas: |
550 L/hr |
MRM conditions: |
Pharmaceuticals listed in Table 1 Pesticides listed in Table 2 |
Since the drinking water samples were free from suspended particulates, they were injected directly into the system. To calculate reproducibility, six replicates of seven different concentrations of each analyte (ranging from 10 ppt to 5,000 ppt) were injected. Overall, the calculated concentrations for 2 orders of magnitude showed excellent reproducibility with CV’s (coefficient of variation) between 0.6 and 14.1%. A quadratic linear regression (0.99) with 1/x2 weight was used for the study. Tables 3 and 4 show the calculated average concentration for six replicates with the corresponding CV values in parenthesis at a 100 ppt spike in drinking water samples.
Figure 2 shows extracted chromatograms of two pesticides (atrazine and carbofuran) in drinking water spiked at 10 ppt. Overall, the peaks show a well-defined Gaussian shape indicating they were refocused well on the analytical column following elution off the SPE extraction column.
The pesticide carbofuran (from the carbamate class of compounds) gave an average RSD (area) of less than 15% across a 3 order of magnitude calibration curve. The baseline signal was flat demonstrating that the 20% methanol wash was effective for removing interferences. Figure 3 shows extracted chromatograms for clotrimazole and tripolidine in drinking water spiked at 10 ppt. Again, the peaks show an excellent Gaussian shape, ensuring a high level of confidence in the results.
The extraction column gave good results after 500 injections of drinking water samples. Figures 4 and 5 show extracted chromatograms at the first, 250th and 500th injections for metolachlor and diphenhydramine with excellent reproducibility on the retention time (CV less than 2%) and a coefficient of variation (CV’s) of less than 15% for signal areas.
The UPLC with on-line SPE Technology using ACQUITY TQD for detection gave an average quantification limit of 10 ppt for pesticides and pharmaceuticals in drinking water samples. The overall reproducibility on the area was less than 15% across 3 orders of magnitude. The extraction column gave excellent results even after 500 injections of drinking water samples.
The UPLC with on-line SPE Technology provides multiple benefits:
The UPLC with On-line SPE System streamlines both the extraction and analysis into a turnkey total solution platform.
720003376, March 2010