This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights the multi-residue analysis of pesticides in oranges using AOAC QuEChERS method by UPLC-MS/MS
|
LC System: |
Waters ACQUITY UPLC System |
|
Column: |
ACQUITY UPLC BEH C18, 2.1 x 100 mm, 1.7 μm |
|
Column Temp: |
40 °C |
|
Sample Temp: |
4 °C |
|
Flow Rate: |
0.3 mL/min. |
|
Mobile Phase A: |
Water + 0.1% formic acid |
|
Mobile Phase B: |
Methanol + 0.1% formic acid |
|
Injection Volume: |
15 μL, Partial loop injection |
|
Time |
Flow Rate |
A% |
B% |
|---|---|---|---|
|
0.00 |
0.3 |
75 |
25 |
|
0.25 |
0.3 |
75 |
25 |
|
7.75 |
0.3 |
5 |
100 |
|
8.50 |
0.3 |
0 |
100 |
|
8.51 |
0.5 |
75 |
25 |
|
10.50 |
0.5 |
75 |
25 |
|
11.00 |
0.3 |
75 |
25 |
|
Instrument: |
Waters ACQUITY TQ Detector |
|
Ionization: |
Positive electrospray (ESI+) |
|
Acquisition: |
Multiple reaction monitoring (MRM) |
720003366, March 2010