This application note demonstrates the ability of the MassTrak AAA Solution to quantify the entire range of amino acids from micro to millimolar levels.
Amino acid analysis of physiological fluids is an important tool used in the study of metabolic pathways. Disturbances to these pathways may give rise to very high levels of one or more amino acids in any given sample while other amino acids in the same sample remain at the low micromolar level or below. It is therefore necessary that the method of analysis offers an analytical measurement range (AMR) that is compatible with the wide range distribution of expected concentrations of amino acids in physiological samples.1
The linear range of current ion exchange methods for amino acid analysis have been previously examined.2,3 However, the reported upper limit of quantification is often equivalent to the concentration of the commercially available physiological standards. While these levels meet the needs of most laboratories, accurate quantification at even higher concentrations must be assured.
In this work, the linear range of the MassTrak AAA Solution (representative chromatogram – Figure 1) is shown to span 1 μM to 10 mM. Accurate quantification of elevated amino acid levels in biological matrices is also demonstrated. These studies show the analytical measurement range of the MassTrak AAA Solution – a range that is comparable to existing methods for amino acid analysis.
Several factors may contribute to imposing an upper limit of quantification in any analytical method. These may include detector linearity, solubility of the analyte, saturating amounts of other sample components, and so on. In any derivatization method, the analysis of elevated levels is limited by the amount of derivatization reagent. MassTrak AAA Reagent must be present in approximately four-fold molar excess for complete derivatization of all amino acids. Therefore the total molar amount of amino acids per derivatization must be considered in defining the AMR. These experiments test both the total amounts of all amino acids, as well as the presence of one amino acid at extremely elevated levels.
Amino acids were prepared at concentrations of 1 μM to 10 mM. At concentrations less than 2.5 mM, amino acids were tested in sets of four. For concentrations above 2.5 mM, amino acids were prepared in smaller subsets. Amino acid samples were derivatized with the MassTrak AAA kit. The amino acids react with 6-aminoquinolyl-N-hydroxysuccinimidyl carbamate (AQC, US Patent 5,296,599. and European Patent EP 0533 200 B1). The protocol followed that described in the MassTrak AAA Solution User’s Guide. Norvaline was used as an internal standard.
Plasma samples were supplemented with elevated concentrations of an amino acid for further studies. Each aliquot of plasma was spiked prior to deproteinization. Norvaline was used as an internal standard and added in the deproteinization step. A typical example:
LC system: |
Waters ACQUITY UPLC System with TUV |
Column: |
MassTrak AAA Column, 2.1 x 150 mm, 1.7 μm |
Column temp.: |
43 °C |
Flow rate: |
400 μL/min. |
Mobile phase A: |
MassTrak AAA Eluent A Concentrate, diluted 1:10 |
Mobile phase B: |
MassTrak AAA Eluent B |
Weak needle wash: |
5/95 Acetonitrile/Water |
Strong needle wash: |
95/5 Acetonitrile/Water |
Gradient: |
MassTrak AAA Standard Gradient |
Detection: |
UV @ 260 nm |
Injection volume: |
1 μL |
Injection mode: |
Partial Loop with Needle Overfill (PLNO) |
In previous work, the MassTrak AAA solution has been shown to exhibit a linear response from 1–2 to 500 μM for the complete amino acid standard.4 However, for each amino acid a linear response over the whole analytical range is required.5 For biological fluids, this corresponds to levels above 500 μM.
In order to accurately evaluate the entire linear range for the MassTrak AAA Solution, certain criteria were set. These included:
All amino acids in the MassTrak AAA standard were analyzed for linear responses over four orders of magnitude: from Lower Limit of Quantification (LLOQ) of 1 μM to an Upper Limit of Quantification (ULOQ) of 10 mM.
For a subset of amino acids, overlaid chromatograms of the complete set of calibration standards are shown in Figure 3. There is no significant distortion of peak shape or retention time that compromises identification of any of the amino acids over the tested range of concentrations. In addition, the detection levels are within the linearity specifications for the ACQUITY TUV Detector. These characteristics indicate that the qualitative properties of the assays are maintained, even over this very wide concentration range.
Empower 2 was used to plot response curves for each of the amino acids. A software generated curve is shown for a representative amino acid (Cit) (Figure 4). The calibration curve is over 4 orders of magnitude with R2 = 0.999, and y intercept = 0.000012. The individual data points in the calibration curve are within 4% of the expected value (Table 2), meeting previously established criteria for linear responses.
Similar results were observed for all of the amino acids in the physiological standard as listed in Table 3. For the majority of amino acids the lower limit of quantification is demonstrated to be 1 μM. However, due to background contamination of some ubiquitous amino acids, such as Gly, an LLOQ of 2 μM is achieved for select analytes.
Various disturbances to amino acid metabolism may give rise to exceptionally high levels of one or more amino acids in physiological fluids. Phenylalanine, for example, has been observed at very high concentrations in biological samples.1 It is important, therefore, to accurately quantify high concentrations of particular amino acids in biological samples.
In this set of experiments, phenylalanine was spiked into pooled human plasma prior to deproteinization. The chromatograms containing increasing levels of spiked phenylalanine are shown in Figure 5. The concentration of phenylalanine in the unspiked or blank plasma was measured and subtracted from the spiked sample analyses. The resulting sample concentrations were within 6% of expected values (Table 4), demonstrating a wide linear range comparable to existing ion-exchange methods.1
In any pre- column derivatization technique, a limiting factor on the upper limit of quantification is the amount of reagent available. A report method was therefore developed to ensure that the MassTrak AAA Solution consistently yields accurate results, even for biological samples containing highly elevated amounts of more than one amino acid.
The MassTrak AAA Solution contains a custom report (Figure 6) in which total amino acid content is measured and compared to the amount of derivatization reagent. If this ratio is sufficient for complete derivatization, the results are acceptable and the sample is “within linear range.” If the levels of total amino acid content are too high for complete derivatization, the report flags the sample for “review” and dilution may be required.
Amino acid analysis is used to study and monitor various metabolic pathways. In a number of disorders, the amino acid concentration can vary widely depending on the physiological state. There is therefore a need to quantify unusually high levels of amino acids in biological samples.
The MassTrak AAA Solution meets these needs, demonstrating a linear response from 1-2 μM up to 10 mM for all the amino acids in the common physiological standard. This method has also been shown to accurately quantify elevated amino acids in biological matrices. However, if a sample’s total amino acid content were to be outside of the method’s linear range, immediate determination can be made with the MassTrak AAA Solution reporting methods.
These results successfully demonstrate the ability of the MassTrak AAA Solution to quantify the entire range of amino acids from micro to millimolar levels. The linear range is comparable to if not greater than that shown for ion exchange methods. More importantly, the linear ranges of all amino acids meet the general guidelines in which linearity must be demonstrated to the highest expected levels.
720003141, July 2009