This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates the LC-MS gradient separation of buprenorphine and buprenorphine glucuronide on XBridge HILIC Columns.
The compounds used in this study are:
|
Column: |
XBridge HILIC, 3.5 μm, 2.1 x 100 mm |
|
Part number: |
186004433 |
|
Mobile phase A: |
10 mM NH4COOH in H2O, 0.125% HCOOH in 50:50 ACN:H2O |
|
Mobile phase B: |
10 mM NH4COOH in H2O, 0.125% HCOOH in 90:10 ACN:H2O |
|
Flow rate: |
0.3 mL/min |
|
Injection volume: |
5 μL |
|
Sample concentration: |
50 ng/mL each |
|
Sample diluent: |
75:25 ACN:MeOH with 0.2% HCOOH |
|
Column temperature: |
30 °C |
|
Instrument: |
Waters ACQUITY UPLC with TQD |
|
MRM: |
Buprenorphine 468.3 > 54.7 Buprenorphine glucuronide 644.3 > 468.3 |
|
Dwell time: |
75 ms |
|
ISD: |
10 ms |
|
ICD: |
10 ms |
|
Time(min) |
Profile |
|
|---|---|---|
|
%A |
%B |
|
|
0.00 |
0.1 |
99.9 |
|
1.05 |
0.1 |
99.9 |
|
4.35 |
99.9 |
0.1 |
|
5.00 |
99.9 |
0.1 |
|
5.01 |
0.1 |
99.9 |
|
6.00 |
0.1 |
99.9 |
WA64077, August 2009