In this study a novel reversed-phase sorbent, Oasis PRiME HLB, is used for highly effective removal of both phospholipids and fats from bovine liver extracts prior to LC-MS/MS analysis. With the new sorbent recoveries of veterinary drugs were similar to results obtained using C18 for clean-up. However, greater than 95% of phospholipids and greater than 85% of fats were effectively removed from the tissue extracts after the simple pass-through SPE procedure.
In order to ensure public health and safety, reliable analytical methods are necessary to determine veterinary drug residue levels in edible tissue samples such as beef liver. The compounds of interest range from highly polar water-soluble compounds to very non-polar fat-soluble compounds. In order to maximize throughput and minimize costs it is desirable to determine the widest possible range of veterinary drug residues in tissue samples with a single analytical method.
Tissue samples, such as bovine muscle and liver, are typically extracted with an acetonitrile based solvent for LC-MS determination of veterinary drug residues. Among the most significant co-extracted substances are fats and polar lipids, particularly phospholipids (lecithin). A gram of bovine liver typically contains about 45 mg of fat, about half the amount usually present in muscle tissue, but still significant. Bovine liver is also a very good source of dietary lecithin (phospholipids); a gram of liver contains about 25 mg of phospholipids, about four times the amount typically found in muscle. Fats can be removed from the acetonitrile based tissue extracts by liquid extraction with hexane or with SPE with octadecyl silica (C18). Although C18 is effective for removal of most non-polar lipids, it does not remove phospholipids. Excessive amounts of phospholipids can shorten LC column life, contribute to ion-suppression, and contaminate the mass spectrometer. In this study a novel reversed-phase sorbent, Oasis PRiME HLB, is used for highly effective removal of both phospholipids and fats from bovine liver extracts prior to LC-MS/MS analysis. With the new sorbent recoveries of veterinary drugs were similar to results obtained using C18 for clean-up. However, greater than 95% of phospholipids and greater than 85% of fats were effectively removed from the tissue extracts after the simple pass-through SPE procedure.
LC system: |
ACQUITY UPLC I-Class with Fixed-Loop Sample Manager |
Column: |
ACQUITY UPLC CSH C18, 1.7 μm, 2.1 mm x 100 mm I.D. |
Mobile phase: |
A: 0.1% formic in water B: 0.1% formic acid in 50:50 acetonitrile/methanol |
Injection vol.: |
7 μL |
Injection mode: |
partial loop injection |
Column temp.: |
30 °C |
Weak needle wash: |
10:90 acetonitrile:water (600 μL) |
Strong needle wash: |
50:30:40 water:acetonitrile:IPA (200 μL) |
Seal wash: |
10:90 acetonitrile: water |
Time (min) |
Flow rate (mL/min) |
%A |
%B |
---|---|---|---|
0.0 |
0.4 |
99.0 |
1.0 |
4.0 |
0.4 |
80.0 |
20.0 |
5.0 |
0.4 |
50.0 |
50.0 |
7.0 |
0.4 |
1.0 |
99.0 |
10.0 |
0.4 |
1.0 |
20.0 |
10.1 |
0.4 |
99.0 |
1.0 |
12.0 |
0.4 |
99.0 |
1.0 |
Mass spectrometer: |
Xevo TQ-XS |
Mode: |
Positive Ion Electrospray |
Source temp.: |
150 °C |
Desolvation temp.: |
400 °C |
Desolvation gas flow: |
1000 L/Hr |
Cone gas flow: |
30 L/Hr |
Collision gas flow: |
0.15 mL/Min |
Data management: |
MassLynx v4.1 |
A 2 g sample of tissue was placed into a 15 mL centrifuge tube containing ceramic homogenizer balls (a Bertin Technologies Precellys Evolution Homogenizer was used for this step). For standards or QC samples the samples were spiked with appropriate amounts of desired analytes. 10 mL 0.2% formic acid in 85:15 acetonitrile/water was added and the samples were homogenized/extracted for 1.5 minutes. The tubes were then centrifuged at 3200 rcf for 5 minutes.
Note: The extraction/precipitation step gives good recovery of most compounds of interest but also extracts significant amounts of fats and phospholipids.
2. Pass-through SPE clean-up:
An Oasis PRiME HLB Cartridge (6 cc, 200 mg) was mounted on a pre-cleaned vacuum manifold. Cartridge conditioning is NOT required, and was NOT performed. The vacuum was set to 2 psi. A 0.6 mL portion of the supernatant was passed-through the Oasis PRiME Cartridge and discarded. Collection tubes were then installed and a 1 mL portion of the supernatant was passed-through the Oasis PRiME Cartridge and collected. A 200 µL aliquot of the pass-through clean-up sample was taken and diluted with 400 µL of 10 mM ammonium formate buffer (pH 4.5) prior to UPLC-MS/MS analysis.
Figure 1 shows the recovery data obtained from replicate analysis of spiked tissue samples (n = 6). Matrix effects averaged about 40%. The chromatograms shown in Figure 2 show the effectiveness of the Oasis PRiME HLB Cartridge for removal of ≥95% of phospholipids from the beef liver extracts. The cartridge also removes more than 90% of hexane extractable fat.
The procedure utilized in this study was developed from methods presented previously.1,2 Although the overall method recoveries averaged above 70 percent, lower recovery was observed for some of the more polar compound classes, such as tetracyclines. Unfortunately, no single solvent extraction step will be highly efficient for all target compounds. For most of the lower recovered compounds the signal response and reproducibility are acceptable for target screening analysis. It is important to understand the contribution of the sample cleanup to any observed recovery losses. The SPE recovery data shown in Figure 3 were obtained from beef liver samples spiked after solvent extraction and prior to SPE clean-up. These data indicate that, for most of the compounds, the Oasis PRiME HLB Cartridge clean-up contributes little to the observed recovery losses. However, for ivermectin, monensin, moxidectin, and novabiocin, the post extraction cleanup did introduce measurable recovery losses. More information on these analytes will be presented in future work.