• Application Note

Multiresidue LC-MS/MS Screening of Veterinary Drugs in Milk Using the Ostro Pass-through Sample Preparation Plate

Multiresidue LC-MS/MS Screening of Veterinary Drugs in Milk Using the Ostro Pass-through Sample Preparation Plate

  • Michael S. Young
  • Kim Van Tran
  • Waters Corporation
Fresh milk bottle and can on table in cowshed

Abstract

This application note describes the use of the Ostro Plate for multi-residue screening for veterinary drugs in milk. Using this plate, milk samples are prepared using a few simple steps to effectively precipitate protein and remove potential interferences.

Benefits

  • Rapid and simple sample preparation for efficient multiclass/multiresidue analysis
  • Removal of potentially interfering phospholipids from prepared samples
  • Fast, sensitive UPLC-MS/MS analysis

Introduction

The Ostro Pass-through Sample Preparation Plate allows for simple and rapid LC-MS determination of basic drugs and related compounds in biological fluids. The Ostro Plate provides a convenient platform for liquid extraction, protein precipitation, filtration, and subsequent removal of phospholipids from the resulting extract. This application note describes the use of the Ostro Plate for multi-residue screening for veterinary drugs in milk. Using this plate, milk samples are prepared using a few simple steps to effectively precipitate protein and remove potential interferences. As the extracted sample is passed through the plate the precipitated proteins are removed by filtration and phospholipid interferences are removed by retention to a proprietary sorbent. After this simple sample preparation, the reconstituted sample is analyzed using UPLC-MS/MS. To demonstrate the suitability of this method, representative compounds were chosen from major classes of veterinary drugs including tetracyclines, fluoroquinolones, sulfonamides, macrolides, beta-lactams, and beta-andrenergics.

Experimental

LC conditions

LC system:

ACQUITY UPLC

Column:

CORTECS UPLC CSH C18+, 1.6 μm, 100 x 2.1 mm

Mobile phase A:

0.1% formic in water

Mobile phase B:

0.1% formic acid acetonitrile

Injection volume:

7 μL

Injection mode:

Partial loop injection

Column temp.:

30 °C

Weak needle wash:

10:90 acetonitrile:water (600µl)

Strong needle wash:

50:30:40 water:acetonitrile:IPA (200µl)

Seal wash:

10:90 acetonitrile:water

Gradient:

Time (min)

Flow (min)

%A

%B

Curve

Initial

0.4

85

15

Initial

2.5

0.4

60

406

6

3.9

0.4

5

95

6

4.9

0.4

5

95

6

5.0

0.4

85

15

6

7.0

0.4

85

15

6

MS conditions

Mass spectrometer:

Xevo TQ MS

Source temp.:

150°C

Desolvation temp.:

500°C

Desolvation gas flow:

1000 L/hr

Cone gas flow:

30 L/hr

Collision gas flow:

0.15 ml/min

Data management:

MassLynxv4.1

summarizes the MRM transitions and instrument parameters
(MRM transition in italics used for quantitation)
Table 1. MRM transitions for this study with associated cone volatage and collision energy.

Sample preparation

Transfer 125 μL milk to a sample well (do not use a sample volume greater than 125 μL). Add 375 μL .2% formic acid in acetonitrile (ACN). Mix well with aspiration. Elute into collection plate well. Add 100 μL 200 mM ammonium formate in 50:50 methanol/ACN and mix well. Evaporate and reconstitute in 100 μL 25:75 ACN/25 mM aqueous ammonium formate. Figure 1 shows a typical setup for SPE using the Ostro Plate.

 

 Typical setup for Ostro Pass-through Plate for analysis of milk
Figure 1. Typical setup for Ostro Pass-through Plate for analysis of milk.

Results and Discussion

Multiresidue Analysis

Compared to a previously published method,1 the Ostro Pass-through Plate procedure gives comparable results. However, the Ostro Plate procedure takes significantly less time and is much more suitable for high throughput analysis. A typical analyst can prepare a batch of ten or more samples in a few minutes compared with a few hours using prior methods. Figure 2 shows a typical LC-MS chromatogram obtained from analysis of a matrix matched standard of enrofloxacin at 10 μg/L. Although performance for most of the other compounds was similar, chlortetracycline, oxytetracycline and lincomycin were not quantifiable at the 10 μg/L level. An alternative method for tetracylines has recently been presented.2 Table 2 shows recovery data observed for multiresidue milk analysis.

Enrofloxacin determined at 10 μg/L in milk
Figure 2. Enrofloxacin determined at 10 μg/L in milk using Ostro Pass-through Plate.
 Recovery data
Table 2. Recovery data (n = 6, bdl = below detection limit).

Phospholipid Removal

Milk contains a significant amount of phospholipids, approximately 1% of the total lipid content.3 These phospholipids can interfere directly with late eluting compounds in LC separations. Also, the buildup of phospholipid residue has been shown to result in poor or deteriorating performance in reversed-phase LC separations.4 Figure 3 shows UPLC-MS/MS analysis of milk samples using an MRM transition (m/z 184>184) chosen for identification of phospholipids. The lower trace shows the result obtained from a milk sample processed using the Ostro Pass-through Plate. The upper trace was obtained using protein precipitation with no SPE cleanup. The Ostro Plate was effective in removal of >98% of phospholipids.

Removal of phospholipids from whole milk
Figure 3. Removal of phospholipids from whole milk (3 % fat) using the Ostro Pass-through Plate.

Conclusion

  •   The Ostro Pass-through Sample Preparation Plate was  demstrated for rapid multiclass extraction of veterinary drugs from milk.
  •   Acceptable performance for screening analysis was observed for all classes of compounds tested except for tetracyclines.
  •   The Ostro Pass-through Sample Preparation Plate was effective for nearly complete removal of phospholipids from milk extracts.

References

  1. Optimized Extraction and Cleanup Protocols for LC-MS/MS Multi-Residue Determination of Veterinary Drugs in Edible Muscle Tissues, Waters Application Note 720004144en
  2. HPLC/UV Determination of Tetracyclines in Milk Using Mixed-Mode SPE and extended Performance [XP] 2.5 μm Columns, Waters Application Note 720004582en
  3. Contarini, G and Povolo, M., Phospholipids in Milk Fat: Composition, Biological and Technological Significance, and Analytical Strategies, Int. J. Mol. Sci, 14(2), 2808–2831(2013)
  4. Eliminating Phospholipids in Drug Discovery Extractions Using a Fast, Generic Sample Clean-up Method, Waters Application Note 720004046en

720005053,  November 2014

Back To Top Back To Top