This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates analysis of food Sugars/Saccharides in honey.
            
        | 
                     Column:  | 
                     ACQUITY UPLC BEH Amide 2.1 x 50 mm, 1.7 μm  | 
            
| 
                     Part Number:  | 
                     186004800  | 
            
| 
                     Mobile Phase A:  | 
                     80/20 MeCN/H2O with 0.2% triethylamine [TEA]  | 
            
| 
                     Mobile Phase B:  | 
                     30/70 MeCN/H2O with 0.2% triethylamine [TEA]  | 
            
| 
                     Flow Rate:  | 
                     0.15 mL/min  | 
            
| 
                     Injection Volume:  | 
                     0.7 μL (PLNO)  | 
            
| 
                     Sample Concentration:  | 
                     Honey and corn syrup at 5-10 mg/mL each  | 
            
| 
                     Sample Diluent:  | 
                     50/50 MeCN/H2O  | 
            
| 
                     Column Temperature:  | 
                     45 °C  | 
            
| 
                     Strong Needle Wash:  | 
                     20/80 MeCN/H2O (800 μL)  | 
            
| 
                     Weak Needle Wash:  | 
                     75/25 MeCN/H2O (500 μL)  | 
            
| 
                     Seal Wash:  | 
                     50/50 MeCN/H2O  | 
            
| 
                     Instrument:  | 
                     Waters ACQUITY UPLC with ELSD  | 
            
| 
                     Gradient:  | 
                     5 minute gradient, 80%-50% MeCN (w/0.2% TEA) with 10 minute re-equilibration  | 
            
| 
                     Time (min)  | 
                     Profile  | 
            |
|---|---|---|
| 
                     A%  | 
                     B%  | 
            |
| 
                     0.00  | 
                     100.00  | 
                     0.00  | 
            
| 
                     5.00  | 
                     40.00  | 
                     60.00  | 
            
| 
                     5.01  | 
                     100.00  | 
                     0.00  | 
            
| 
                     15.00  | 
                     100.00  | 
                     0.00  | 
            
| 
                     Gain:  | 
                     200  | 
            
| 
                     Pressure:  | 
                     40 psi  | 
            
| 
                     Drift Tube Temperature:  | 
                     40 °C  | 
            
| 
                     Nebulizer:  | 
                     Cooling  | 
            
| 
                     Data Rate:  | 
                     10 pps  | 
            
| 
                     Filter Time Constant:  | 
                     Normal  | 
            
            
        WA60124, October 2009