This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates analysis of ecstasy (MDMA) and metabolites.
The compounds used in this study are –
|
Column: |
Xterra MS C18 2.1 x 150 mm, 3.5 μm |
|
Part number: |
186000408 |
|
Mobile phase A: |
20 mM NH4HCO3, pH 9.0 |
|
Mobile phase B: |
MeOH |
|
Flow rate: |
0.2 mL/min |
|
Isocratic mobile phase composition: |
70% A; 30% B |
|
Injection volume: |
15 μL |
|
Temperature: |
30 °C |
|
Detection: |
MS ESI+ |
|
Instrument: |
Alliance 2695, Micromass ZQ |
|
Micromass ZQ ESI+ |
|
|---|---|
|
Capillary (KV): |
3.0 |
|
Cone (V): |
30 |
|
Extractor: |
3.0 |
|
RF Lens: |
0.5 |
|
Source temp.: |
150 |
|
Desolvation temp.: |
350 |
|
Cone gas flow (L/Hr): |
50 |
|
Desolvation gas flow (L/Hr): |
500 |
|
LM resolution: |
15 |
|
HM resolution: |
15 |
|
Ion energy: |
0.1 |
|
Multiplier (V): |
650 |
WA20738.047, June 2002