This application note shows the newly developed Waters Micromass Quattro Premier Tandem Quadrupole Mass Spectrometer for the analysis of 17- hydroxyprogesterone in protein-precipitated plasma.
The Quattro Premier can be used for quantitation over five orders of linear dynamic range
Atmospheric Pressure Chemical Ionization (APCI) is often cited as the technique of choice for improved linearity in quantitative LC-MS/MS. In this application note we show the newly developed Waters Micromass Quattro Premier Tandem Quadrupole Mass Spectrometer in the analysis of 17- hydroxyprogesterone in protein-precipitated plasma. The calibration curve was plotted over five orders linear dynamic range of concentration, indicating excellent dynamic range that will be beneficial in dose ranging studies.
A standard solution of 17-hydroxyprogesterone was prepared in methanol (1 mg/mL). Subsequent dilutions were made in human plasma to generate a calibration curve from 1–100,000 pg/μL (5–500,000 pg on column). The calibration standards were then protein-precipitated by adding acetonitrile (1:1), the resultant mixtures were centrifuged (ca 3000 rpm, 10 mins) and the supernatant taken for analysis by LC-MS/MS. Duplicate injections were made of each standard.
LC system: |
Waters 1525μ binary HPLC pump |
Sample manager: |
Waters 2777 |
Column: |
Waters Symmetry C18 4.6 x 50 mm, 3.5 μm |
Solvents: |
A - 70% water 30% MeOH 10 mM ammonium acetate 0.005% acetic acid B - 100% MeOH 10 mM ammonium acetate 0.005% acetic acid |
Injection volume: |
5 μL |
Mass spectrometer: |
Waters Micromass Quattro Premier |
Ion mode: |
APCI +ve |
Corona: |
5 mA |
Cone voltage: |
40 V |
Collision energy: |
18 eV |
Detection mode: |
MRM (331.3 > 108.9) |
Dwell: |
0.3 seconds |
Collision gas: |
Argon (2.7 x 10-3 mbar) |
The plot of peak area against concentration showed good linearity over the range 5–500,000 pg on column. The calibration line was plotted using a linear fit with 1/x2 weighting and gave a correlation coefficient of >0.99 (Figure 1).
Figure 2 shows the QuanLynx results table demonstrating that all the calibration points gave back-calculated values within ±6% of the theoretical concentrations at all concentrations except for the LLOQ which were within ±13%.
The new Waters Quattro Premier Tandem Quadrupole Mass Spectrometer has been developed for quantitative LC-MS/MS. The results also show that the Quattro Premier can be used for quantitation over five orders of linear dynamic range, i.e. 5 to 500,000 pg on column for 17-hydroxyprogesterone.
720000825, March 2004