This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights the analysis of nucleotide triphosphates using XTerra MS C18 columns.
Compounds used in this study are:
1. Deoxycytidine 5’-triphosphate
2. Thymidine 5’-triphosphate
3. Deoxyguanosine 5’-triphosphate
4. Adenosine 5’-triphosphate
            
        | 
                     Column:  | 
                     XTerra MS C18 4.6 x 30 mm, 2.5 μm (p/n: 186000600)  | 
            
| 
                     Mobile phase A:  | 
                     50 mM Triethylammonium acetate, pH 7.0, 2 mM EDTA  | 
            
| 
                     Mobile phase B:  | 
                     ACN  | 
            
| 
                     Flow rate:  | 
                     1.0 mL/min  | 
            
| 
                     Injection volume:  | 
                     10 μL  | 
            
| 
                     Detection:  | 
                     UV @ 260 nm  | 
            
| 
                     Temperature:  | 
                     20 °C  | 
            
| 
                     Instrument:  | 
                     Alliance 2695, 2996 PDA  | 
            
            
        
            
        WA20738.081, June 2002