This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights the analysis of aromatic amino acids.
Column: |
Atlantis T3, 4.6 x 50 mm, 3 μm |
Part Number: |
186003726 |
Mobile Phase A: |
H2O |
Mobile Phase B: |
MeOH |
Mobile Phase C: |
100 mM HCOONH4, pH 3.0 |
Flow Rate: |
1.0 mL/min |
Injection Volume: |
10 μL |
Sample: |
Tyrosine (150 μg/ mL), Phenylalanine (200 μg/mL), Trytophan (30 μg/mL) in H2O |
Column Temperature: |
30 ˚C |
Sample Temperature: |
20 ˚C |
Detection: |
UV @ 254 nm |
Sampling Rate: |
5 point/sec |
Filter Response: |
0.2 |
Instrument: |
Waters Alliance 2695 with 2996 PDA |
Time (min) |
Profile |
||
---|---|---|---|
%A |
%B |
%C |
|
0.00 |
90 |
0 |
10 |
10.00 |
10 |
80 |
10 |
11.00 |
90 |
0 |
10 |
15.00 |
90 |
0 |
10 |
WA60221, January 2009