This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights the analysis of statins using XBridge Phenyl columns.
|
Columns: |
XBridge Phenyl, 4.6 x 100 mm, 3.5 μm p/n: 186003334 |
|
Mobile phase A: |
H2O |
|
Mobile phase B: |
ACN |
|
Mobile phase C: |
100 mM NH4HCO3 |
|
Flow rate: |
1.2 mL/min |
|
Injection volume: |
20 μL |
|
Sample: |
Atorvastatin (10 μg/mL), Simvastatin (10 μg/mL), Lovastatin (10 μg/mL), Pravastatin Sodium (10 μg/mL), in H2O/NH4HCO3 (90/10) |
|
Column temp.: |
25 °C |
|
Sample temp.: |
15 °C |
|
Detection: |
UV @ 248 nm |
|
Sampling Rate: |
5 points/sec |
|
Filter Response: |
0.2 |
|
Instrument: |
Alliance 2695 with 2996 PDA |
|
Time (min) |
%A |
%B |
%C |
|---|---|---|---|
|
0 |
65 |
15 |
20 |
|
20 |
30 |
50 |
20 |
|
21 |
65 |
15 |
20 |
|
25 |
65 |
15 |
20 |
WA60189, June 2007