This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights the analysis of paclitaxel impurities.
Paclitaxel (taxol) has been studied in this application brief.
            
        | 
                     Column:  | 
                     Symmetry C18, 4.6 x 100 mm, 3.5 μm (p/n: WAT066220)  | 
            
| 
                     Mobile phase:  | 
                     Water/acetonitrile/methanol 57:34:9  | 
            
| 
                     Flow rate:  | 
                     1.4 mL/min  | 
            
| 
                     Injection volume:  | 
                     25 μL of paclitaxel, 1 mg/mL in methanol  | 
            
| 
                     Detection:  | 
                     UV @ 230 nm  | 
            
            
        WA31763.117, June 2003