Analysis of Aldosterone in Plasma for Clinical Research using the ACQUITY UPLC I-Class/Xevo TQ-S micro System
This is an Application Brief and does not contain a detailed Experimental section.
For research use only. Not for use in diagnostic procedures.
Abstract
To demonstrate the capability of the ACQUITY™ UPLC™ I-Class System and Xevo™ TQ-S micro Mass Spectrometer to quantify low levels of aldosterone in plasma for clinical research using a highly selective sample preparation technique.
Benefits
- Analytically sensitive analysis of aldosterone in plasma (low calibrator 42 pmol/L).
- Incorporation of automated sample preparation optimizes analytical sensitivity, reduces sample handling time, and alleviates the potential for operator error.
- Total precision ≤7.2% across the calibration range for plasma QC samples over five days (n=25).
Introduction
Aldosterone is a mineralocorticoid steroid hormone, that is assessed in clinical research studies to help understand the pharmacological mechanism of aldosterone synthase inhibitors (ASIs).1 Circulating levels of aldosterone in blood are typically found at low concentrations (<100 pmol/L), which makes its analysis particularly challenging. Successfully quantifying these low levels necessitate the use of a mass spectrometer with high analytical sensitivity in conjunction with highly selective sample preparation techniques.
Here we introduce the Xevo TQ-S micro utilizing innovative technology such as the Stepwave™ ion source technology to improve method robustness and reduce background noise, which enables accurate and precise quantification of low level analytes such as aldosterone.
Experimental
The Solution
In this technology brief, a UPLC-MS/MS method for the analysis of plasma aldosterone was successfully employed using the Xevo TQ-S micro in conjunction with automated selective solid phase extraction sample preparation. The sample preparation was automated on a Tecan® Freedom EVO100/4 using 96-well Oasis™ MAX μElution SPE plates followed by analysis on the MassLynx™ 4.1 controlled UPLC-MS/MS as outlined in application note 720005262.

Sample Preparation and UPLC-MS/MS Analysis
Using the Tecan Freedom EVO100/4, plasma samples were diluted with internal standard, zinc sulphate, methanol, and phosphoric acid. Following centrifugation, sample supernatant was loaded onto the 96-well Oasis MAX μElution SPE plate (p/n:186001829) following conditioning and equilibration. Consecutive washes with phosphoric acid, ammonia in 10% methanol, and water were performed. Samples were eluted with 70% aqueous methanol followed by water.
30 μL of each extracted sample was injected on an ACQUITY UPLC I-Class System/Xevo TQ-S micro System utilizing a water/methanol gradient and a CORTECS™ UPLC C18 Column (p/n: 186007095). The MRM parameters used in this analysis are shown in Table 1.


Results and Discussion
Calibration curves, quality control (QC) samples, and plasma samples were extracted and analyzed over five separate days. The S/N for the low calibrator at 42 pmol/L was >25:1 over the five days, achieved through low background noise on the instrument, selectivity of the MRM trace and a clean SPE sample extract. In addition, the correlation coefficient demonstrates excellent linearity across a range of 42–4161 pmol/L.
Note: To convert SI units to conventional mass units divide by 2.774 for aldosterone (pmol/L to pg/mL).
The additional selectivity provided by the Oasis MAX chemistry provides a clean sample extract for ESI MS analysis of aldosterone. This is observed in the extraction and quantification of 49 pmol/L of aldosterone in plasma using the Xevo TQ-S micro (Figure 1).
Reproducibility of the method was assessed by extracting and quantifying plasma samples using six replicates at low (128 pmol/L), mid (1011 pmol/L), and high (2926 pmol/L) concentrations. All results were ≤7.2% RSD as shown in Table 3.


Conclusion
A UPLC-MS/MS method for the analysis of plasma aldosterone for clinical research has been developed using an ACQUITY UPLC I-Class/Xevo TQ-S micro System. This method incorporates highly efficient automated sample preparation using a Freedom EVO 100/4 and 96-well Oasis MAX μElution SPE plates. The robust and analytically sensitive Xevo TQ-S micro has been shown to provide excellent precision and linearity of response across all plasma aldosterone levels that were tested.
References
- Schumacher CD et al. Aldosterone Synthase Inhibition for the Treatment of Hypertension and the Derived Mechanistic Requirements for a New Therapeutic Strategy. J Hypertens. Oct 2013; 31(10): 2085–2093.
720006313, June 2018