        This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights Oasis 2x4 Method which is a viable approach to SPE sorbent and protocol selection.
In order to prove that the Oasis 2x4 Method is a viable approach to SPE sorbent and protocol selection, a group of representative small molecules was spiked into rat plasma. The molecules include a base, a quaternary ammonium salt, a neutral, an acid and a strong acid. The spiked plasma samples were then extracted following the Oasis 2x4 Method. All molecules elute in the correct Elute 1 or Elute 2 fractions according to what theory predicted, proving that the method is viable.
Follow the simple steps outlined in this flow chart to achieve high recoveries and the cleanest extracts:
 
            
        10 mg Oasis 96-well Plates
| Condition: | 500 μL MeOH | 
| Equilibrate: | 500 μL H2O | 
| Load: | 500 μL sample (250 μL plasma diluted 1:1 with 4% H3PO4 in H2O) | 
| Wash 1: | 500 μL 2% HCOOH in H2O | 
| Elution 1: | 2 x 125 μL MeOH | 
| Elution 2: | 2 x 125 μL 5% NH4OH in MeOH | 
| Dilution: | 250 μL water | 
| Condition: | 500 μL MeOH | 
| Equilibrate: | 500 μL H2O | 
| Load: | 500 μL sample (250 μL plasma diluted 1:1 with 4% H3PO4 in H2O) | 
| Wash 1: | 5% NH4OH in H2O | 
| Elution 1: | 2 x 125 μL MeOH | 
| Elution 2: | 2 x 125 μL 2% HCOOH in MeOH | 
| Dilution: | 5% NH4OH in H2O (To neutralize acid for high pH LC) | 
| Column: | XBridge C18 2.1 x 20 mm IS, 3.5 μm | 
| Mobile Phase A: | 10 mM NH4HCO3, pH 10 | 
| Mobile Phase B: | 10 mM NH4H | 
| Injection Volume: | 10.0 μ L | 
| Column Temperature: | Ambient | 
| Detection: | UV @ 254 nm (Prednisone) | 
| Instrumentation: | 2777 Sample Manager, 1525μ Binary HPLC Pump, Quattro Premier and 2996 PDA | 
| Time (min) | Profile | |
|---|---|---|
| %A | %B | |
| 0 | 95 | 5 | 
| 3.0 | 5 | 95 | 
| 4.8 | 5 | 95 | 
| 5.0 | 95 | 5 | 
| 7.0 | 95 | 5 | 
| Quattro Premier | |
|---|---|
| Capillary: | 3.4 kV | 
| Source Temp.: | 120 °C | 
| Desolvation Temp.: | 350 °C | 
| Cone Gas Flow: | 50 L /Hr | 
| Desolvation Gas Flow: | 700 L /Hr | 
| Collision Cell Pressure: | 2.59e-3 mbar | 
| MRM Transitions: | Imipramine 281.2 > 85.95 ESI+ Decanesulfonic Acid 220.9 > 79.7 ESI - Ibuprofen 205.2 > 161 ESI - Valethamate 306.3 > 162.8 ESI+ | 
 
            
         
            
         
            
         
            
        WA60090, June 2007