This is an Application Brief and does not contain a detailed Experimental section.
The resolving power of the Waters Micromass MALDI micro MX is demonstrated in reflectron and linear mode, in positive ion and negative ion modes.
In this test the MALDI micro MX always surpassed the instrument specifications
Mass resolution is the ability to separate two adjacent masses. It defines the performance of an analyzer and is a measure of the extent to which adjacent ions can be distinguished from each other with less than a defined amount of overlap. For a single peak this can be calculated from the width of a peak at a defined height and is usually referred to as Full Width Half Maximum (FWHM). FWHM is the resolution measured with a 50% valley overlap between peaks of adjacent m/z.
Resolution can be calculated using the following equation:
where m is the observed m/z and Δm is the measured width at half peak height. Two key advantages of increasing resolution are:
It is often assumed that high resolution is always required for good mass accuracy, this has however been shown not to be the case.1
The resolving power of the Waters Micromass MALDI micro MX is demonstrated in reflectron and linear mode, in positive ion and negative ion modes.
Sample: Adrenocorticotropic hormone clip 18–39 (ACTH) at a concentration of 2 pmol/μL in 0.1% trifluoroacetic acid (TFA).
Matrix: Purified α-cyano-4-hydroxy cinnamic acid matrix was obtained from Waters (186002331). Matrix was prepared at 2 mg/mL in 1:1 acetonitrile:aqueous 0.1% TFA.
Samples were mixed 1:1 with matrix and 1 μL was spotted directly onto the target plate and allowed to air dry prior to analysis.
Data acquisition: Six separate acquisitions in positive and negative mode.
Sample: Cytochrome C (equine) at a concentration of 2 pmol/μL in 0.1% TFA.
Matrix: The matrix was prepared and used in 2 stages,
1 μL of Matrix 1 was applied to the target plate and allowed to air dry then Matrix 2 was mixed 1:1 with the sample and 1 μL was spotted on top of the Matrix 1 thin film.
Data acquisition: Six separate acquisitions in positive and negative mode.
The results obtained are summarised in Table 1. Examples for each mode are shown in Figures 2–5. The MALDI micro MX exceeded the specified resolution in every mode.
720001151, April 2005