This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates ß2-Agonists in pork and pig liver tissues.
            
        | 
                     Instrument:  | 
                     Waters Alliance HPLC 2695 System  | 
            
| 
                     Analytical column:  | 
                     Atlantis dC18, 2.1 x 150 mm, 5 μm  | 
            
| 
                     Guard column:  | 
                     Atlantis dC18, 2.1 x 10 mm, 5 μm  | 
            
| 
                     Flow rate:  | 
                     0.2 mL/min  | 
            
| 
                     Mobile phase:  | 
                     A. 0.1% formic acid B. 0.1% formic acid in acetonitrile  | 
            
| 
                     Injection volume:  | 
                     20 μL  | 
            
| 
                     Column temperature:  | 
                     35 °C  | 
            
| 
                     Time (min)  | 
                     A%  | 
                     B%  | 
            
|---|---|---|
| 
                     0  | 
                     96  | 
                     4  | 
            
| 
                     2  | 
                     96  | 
                     4  | 
            
| 
                     8  | 
                     77  | 
                     23  | 
            
| 
                     21  | 
                     77  | 
                     23  | 
            
| 
                     22  | 
                     5  | 
                     95  | 
            
| 
                     25  | 
                     5  | 
                     95  | 
            
| 
                     25.5  | 
                     96  | 
                     4  | 
            
| 
                     Instrument:  | 
                     Waters Quattro Premier  | 
            
| 
                     Ionization mode:  | 
                     Positive electrospray (ESI+) Multiple reaction monitoring  | 
            
            
        
            
        
            
        Pig liver tissues were spiked with 11 β2-agonists standard mixture of concentrations 0.5 ng/g, 1 ng/g and 2 ng/g respectively. The SPE recoveries are between 89.4% and 110.5%, RSD are between 1% and 2.8%.
720002601, April 2008