This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights the analysis of antibacterials using XBridge Phenyl columns.
Compounds used in this study are: 1. Carbadox 2. Thiamphenicol 3. Furazolidone 4. Oxolinic Acid 5. Sulfadimethoxine 6. Sulfaquinoxaline 7. Nalidixic Acid.
|
Columns: |
XBridge Phenyl, 4.6 x 100 mm, 3.5 μm p/n: 186003334 |
|
Mobile phase A: |
20 mM KH2PO4, pH 2.5 |
|
Mobile phase B: |
ACN |
|
Flow rate: |
1.0 mL/min |
|
Injection volume: |
10 μL |
|
Sample: |
Carbadox (10 μg/mL), Thiamphenicol (100 μg/mL), Furazolidone (10 μg/mL), Oxolinic Acid (10 μg/mL), Sulfadimethoxine (10 μg/mL), Sulfaquinoxaline (10 μg/mL), Nalidixic Acid (10 μg/mL) in KH2PO4/ACN (80/20) |
|
Column temp.: |
30 °C |
|
Sampling temp.: |
15 °C |
|
Detection: |
UV @ 254 nm |
|
Sampling Rate: |
5 points/sec |
|
Filter Response: |
0.2 |
|
Instrument: |
Alliance 2695 with 2996 PDA |
|
Time(min) |
%A |
%B |
|---|---|---|
|
0 |
80 |
20 |
|
2 |
80 |
20 |
|
7 |
75 |
25 |
|
15 |
70 |
30 |
|
16 |
80 |
20 |
|
20 |
80 |
20 |
WA60191, June 2007