For forensic toxicology use only.
This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights the analysis of methamphetamines using XTerra Columns.
The compounds analyzed in this study are:
            
        
            
        | 
                     Column:  | 
                     XTerra RP18 3.0 x 100 mm, 3.5 µm  | 
            
| 
                     Part number:  | 
                     186000420  | 
            
| 
                     Mobile phase A:  | 
                     5 mM NH4HCO3, pH 9.6  | 
            
| 
                     Mobile phase B:  | 
                     ACN  | 
            
| 
                     Flow rate:  | 
                     0.6 mL/min  | 
            
| 
                     Isocratic mobile phase composition:  | 
                     80%A; 20%B  | 
            
| 
                     Injection volume:  | 
                     5 µL  | 
            
| 
                     Sample concentration:  | 
                     40 pg/µL  | 
            
| 
                     Temperature:  | 
                     40 °C  | 
            
| 
                     Detection:  | 
                     Fluorescence 280ex 326em Gain 100 TC = 1s  | 
            
| 
                     Instrument:  | 
                     Alliance 2690, 2475  | 
            
| 
                     Compounds  | 
                     USP Tailing  | 
            
|---|---|
| 
                     3,4-methylenedioxyamphetamine (MDA)  | 
                     0.95  | 
            
| 
                     3,4-methylenedioxymethamphetamine (MDMA)  | 
                     1.10  | 
            
            
        WA20738.065, June 2002