• Application Note

Determination of Highly Polar Cationic Pesticides and Plant Growth Regulators in Food Using UPLC-MS/MS

Determination of Highly Polar Cationic Pesticides and Plant Growth Regulators in Food Using UPLC-MS/MS

  • Benjamin Wuyts
  • Janitha De-Alwis
  • Euan Ross
  • Simon Hird
  • Waters Corporation
Farmer spraying wheat field

Abstract

In this application note, example performance data is provided from Waters ACQUITY UPLC I-Class System and Xevo TQ-S micro on three commodities which represent high water content and high starch, low water content sample types. The aim of this study was to evaluate the combination of the ACQUITY UPLC I-Class System coupled with Xevo TQ-S micro for the determination of highly polar cationic pesticide residues and plant growth regulators in several food commodities.

Benefits

Provides a direct, single extraction LC-MS/MS method for the analysis of various highly polar cationic pesticides and plant growth regulators in cereals, fruit, and vegetable commodities.

Introduction

The European Union Reference Laboratory for Pesticides Single Residue Methods (EURL-SRM) published the QuPPe (Quick Polar Pesticides)1 methods for the simultaneous analysis of a number of highly polar pesticides. To meet the needs of analyzing highly polar pesticides by LC-MS/MS, details on a number of chromatographic methods have been provided including one based upon hydrophilic interaction liquid chromatography (HILIC)1-2 for the determination of a series of cationic and polar basic analytes.

Although several compounds included in this application note are approved for use in Europe (maximum residue limits (MRLs) are listed in Table 1), other pesticide/crop combinations are not and default MRLs apply.3 As well as being a separate contaminant of interest, melamine4 is also a metabolite of cyromazine, although it is not yet part of the residue definition used for enforcement purposes.  

Table 1. Current MRLs2-3 in the three representative matrices for the compounds included in this application note.
Table 1. Current MRLs2-3 in the three representative matrices for the compounds included in this application note. 

In this application note, example performance data is provided from Waters ACQUITY UPLC I-Class System and Xevo TQ-S micro on three commodities which represent high water content and high starch, low water content sample types. Organic wheat flour, cucumber, and apple, were extracted following the QuPPe method,1 to assess various performance factors of the UPLC-MS/MS method such as calibration linearity, retention time stability, method precision, and trueness. 

Experimental

Sample preparation and extraction

Homogenized organic apple and cucumber were extracted using the QuPPe method1 as shown in Figure 1. For wheat flour only 5 g of sample was taken and 10 mL of LCMS grade water was added to this before extraction with the acidified methanol. Before the centrifugation step, the wheat flour was placed in a freezer at -20 °C for 2 hrs. The supernatant from the QuPPe extracts were then filtered using a 0.45 µm PVDF filter, spiked with the pesticide mix and analyzed using the liquid chromatography, mass spectrometry method highlighted below. Method performance information for analyte recovery can be found in the QuPPe document.1

QuPPe sample extraction workflow.
Figure 1. QuPPe sample extraction workflow for organic apple and cucumber.  

The performance of the LC-MS/MS step of the method was assessed using SANTE guidelines.5 Solutions of matrix-matched standards were prepared over the range 0.002 to 0.200 mg/kg (1.0 to 100.0 ng/mL in vial concentration) in apple and cucumber, 0.004 to 0.400 mg/kg (1.0 to 100.0 ng/mL in vial concentration) in wheat flour. Replicate injections at two concentration levels were run between bracketed calibration curves to assess the performance of the method. No isotopically labelled standards were used for this analysis.

UPLC conditions

UPLC system:

ACQUITY UPLC I-Class with fixed-loop Sample Manager

Column:

ACQUITY UPLC BEH Amide, 1.7 μm, 2.1 × 100 mm (p/n: 186004801)

Mobile phase A:

50 mM Ammonium formate (pH 2.9, adjusted with LCMS grade formic acid)

Mobile phase B:

Acetonitrile

Flow rate:

0.5 mL/min

Injection volume:

0.5 μL (partial loop needle overfill)

Weak wash solvent:

90:10 acetonitrile:water

Strong wash solvent:

10:90 acetonitrile:water

Column temp.:

40 °C

Sample temp.:

10 °C

Run time:

10 min

Gradient:

Time (min)

%A

%B

Curve

0.00

3.0

97.0

Initial

0.50

3.0

97.0

6

4.00

30.0

70.0

6

5.00

40.0

60.0

6

6.00

40.0

60.0

6

6.10

3.0

97.0

6

10.00

3.0

97.0

6

MS conditions

MS system:

Xevo TQ-S micro

Ionization:

ESI+

Capillary voltage:

0.5 kV

Desolvation temp.:

600 °C

Desolvation gas flow:

1000 L/Hr

Source temp.:

150 °C

Cone gas flow:

150 L/Hr

Nebulizer gas pressure:

7 Bar

MRM transitions:

MRM transitions
* denotes transition used for quantification.

Results and Discussion

The method was found to give good retention for all compounds, greater than two times the column void volume, as indicated in the SANTE guidelines.5 Overall the method provided acceptable separation and excellent peak shapes for all compounds. Example chromatograms for chlormequat, mepiquat, and propamocarb in matrix at the 1 ng/mL in vial concentration level, are shown in Figure 2. Retention time stability was also assessed according to the SANTE guidelines (±0.1 min),5 retention times within and between matrices were within 0.1 min for all compounds. Retention times for each compound in the representative matrices can be found in Figure 3.

Figure 2. Example chromatograms for A. chlormequat in cucumber, B. mepiquat in apple, and C. propamocarb in wheat flour at 1.0 ng/mL in vial concentration. Excellent sensitivity for both MRM transitions for each compound was achieved with only a 0.5 µL injection volume.
Figure 2. Example chromatograms for A. chlormequat in cucumber, B. mepiquat in apple, and C. propamocarb in wheat flour at 1.0 ng/mL in vial concentration. Excellent sensitivity for both MRM transitions for each compound was achieved with only a 0.5 µL injection volume.
Bracketed matrix-matched calibration curves.
Figure 3. Bracketed matrix-matched calibration curves for cucumber and wheat flour for chlomequat, 1 ng/mL to 100 ng/mL in vial concentration.

The linearity of the method was assessed using bracketed matrix-matched calibration curves for each matrix, without the use of labeled internal standards; Figure 3 shows the calibration curves for chlormequat in organic cucumber and wheat flour. The linearity of response and calibration range in the tested matrices for all compounds assessed in this study, are shown in Table 2. The concentration levels take into account that only 5 g of wheat flour was taken for extraction. All compounds gave excellent  linear response and residuals (back calculated concentrations) were within the 20% tolerance of the SANTE guidelines.5

Matrix-matched calibration linearity of response and calibration range.
Table 2. Matrix-matched calibration linearity of response and calibration range (mg/kg) for each compound in the three tested matrices. The calibration range for wheat flour takes into account that only 5 g of sample is used for the extraction.

Replicate (n=15) injections were run for two levels, 5.0 ng/mL and 20.0 ng/mL in vial concentrations. The calculated mean concentrations and precision for the tested compounds in all three matrices can be seen in Figure 4. Excellent accuracy and precision was achieved for all compounds, within 15% of the target concentration value and %RSD below 5%.

Ion ratios of the replicate injections agreed well with expected reference values and all were within the required tolerance5 (±30%).  An example of the ion ratios given by each of the (n=15) replicate levels for chlomequat in wheat flour can be seen in Figure 5. 

Replicate injection data.
Figure 4. Replicate injection data for all tested compounds in apple, cucumber, and wheat flour. The primary axis is the mean %trueness to the target in vial concentration level and the secondary axis is the %RSD at each level (n=15).
 Calculated ion ratios.
Figure 5. Calculated ion ratios (target/quan) for chlormequat at each replicate level (n=15) in wheat flour. 

Conclusion

The aim of this study was to evaluate the combination of the ACQUITY UPLC I-Class System coupled with Xevo TQ-S micro for the determination of highly polar cationic pesticide residues and plant growth regulators in several food commodities. The Xevo TQ-S micro provided excellent, fit-for-purpose performance in terms of sensitivity, linearity, and calibration range for all of the tested matrices. The trueness and precision of this UPLC-MS/MS method determined at two matrix QC levels with 15 replicate injections was found to be acceptable for all compounds. Overall the performance data indicate that the configuration of the ACQUITY UPLC I-Class coupled with Xevo TQ-S micro, when used in combination with the ACQUITY UPLC BEH Amide Column and an established extraction protocol such as QuPPe, is suitable for checking MRL/tolerance compliance in routine laboratory testing for these target compounds. 

References

  1. QuPPe Method. European Commission (2017). http://www.eurl-pesticides.eu/userfiles/file/EurlSRM/meth_QuPPe-PO_EurlSRM.pdf (Accessed online 7 February 2018).
  2. A Comprehensive Guide to HILIC: Hydrophillic Interaction Chromatography. (2014) Waters Corporation (p/n: 715002531). (Accessed online 30 August 2018).
  3. EU Pesticide Database. European Commission (2016) [Online] http://ec.europa.eu/food/plant/pesticides/eu-pesticidesdatabase/public/?event=pesticide.residue.selection&language=EN.
  4. International Experts Limit Melamine Levels in Food. World Health Organisation (2010) http://www.who.int/mediacentre/news/releases/2010/melamine_food_20100706/en/ (Accessed online 30 August 2018).
  5. Document No. SANTE 11813/2017. Guidance Document on Analytical Quality Control and Method Validation Procedures for Pesticides Residues Analysis in Food and Feed European Union (2017).

720006410, October 2018

Back To Top Back To Top