This is an Application Brief and does not contain a detailed Experimental section.
This application brief describes UPLC-MS analysis of food sugars/saccharides in beer.
Compounds analysed in beer are:
            
        | 
                     Column:  | 
                     ACQUITY UPLC BEH Amide 2.1 x 100 mm, 1.7 μm  | 
            
| 
                     Part Number:  | 
                     186004801  | 
            
| 
                     Mobile Phase A:  | 
                     80/20 MeCN/H2O with 0.1% ammonium hydroxide [NH4OH]  | 
            
| 
                     Mobile Phase B:  | 
                     30/70 MeCN/H2O with 0.1% ammonium hydroxide [NH4OH]  | 
            
| 
                     Flow Rate:  | 
                     0.13 mL/min  | 
            
| 
                     Gradient:  | 
                     10 minute gradient, 75%-45% MeCN (w/0.1% NH4OH) with 25 minute re-equilibration  | 
            
| 
                     Injection Volume:  | 
                     2.0 μL (PLNO)  | 
            
| 
                     Sample Concentration:  | 
                     Standards at 10 μg/mL, Beer at 50% dilution  | 
            
| 
                     Sample Diluent:  | 
                     50/50 MeCN/H2O  | 
            
| 
                     Column Temperature:  | 
                     35 °C  | 
            
| 
                     Strong Needle Wash:  | 
                     20/80 MeCN/H2O (800 μL)  | 
            
| 
                     Weak Needle Wash:  | 
                     75/2MeCN/H2O 5 (500 μL)  | 
            
| 
                     Seal Wash:  | 
                     50/50 MeCN/H2O  | 
            
| 
                     Instrument:  | 
                     Waters ACQUITY UPLC with SQ  | 
            
| 
                     Time(min)  | 
                     Profile  | 
            |
|---|---|---|
| 
                     %A  | 
                     %B  | 
            |
| 
                     0.00  | 
                     90.00  | 
                     10.00  | 
            
| 
                     10.00  | 
                     30.00  | 
                     70.00  | 
            
| 
                     10.01  | 
                     90.00  | 
                     10.00  | 
            
| 
                     35.00  | 
                     90.00  | 
                     10.00  | 
            
| 
                     Ionization Mode:  | 
                     ES-  | 
            
| 
                     Capillary:  | 
                     2.8 kV  | 
            
| 
                     Cone Voltage:  | 
                     25 V (fructose, glucose, maltotriose); 40V (sucrose and maltose)  | 
            
| 
                     Source Temp:  | 
                     120 °C  | 
            
| 
                     Desolvation Temp:  | 
                     350 °C  | 
            
| 
                     Desolvation Gas Flow:  | 
                     500 L/Hr  | 
            
| 
                     Cone:  | 
                     50 L/Hr  | 
            
| 
                     SIR (m/z):  | 
                     179.0 (fructose, glucose); 341.1 (sucrose, maltose); 503.2 (maltotriose)  | 
            
| 
                     Dwell Time:  | 
                     0.04 s  | 
            
            
        WA60126, October 2009