This application note demonstrate the dramatic improvement that a new technology, UltraPerformance LC (UPLC), brings to pharmaceutical science of metabolite ID by exploring the metabolites of acetaminophen in human urine.
Many of us are familiar with the popular over-the-counter headache medicine acetaminophen, but have you ever wondered what happens to this compound after it enters the humanbody? In simple terms, most drugs are converted into other compounds, metabolites, by enzymatic action in the body. These metabolites may or may not have an adverse effect on one’s health, and it is therefore crucial to study of the metabolic fate of drugs. For example, an overdose of acetaminophen may generate toxic metabolites that could lead to adverse effects. The prediction of these toxicological events is an ever-present challenge in the pharmaceutical industry. These toxic events are normally characterized by the presence of certain metabolites such as glutathione. High Performance Liquid Chromatography/Mass Spectrometry (HPLC-MS) has been routinely used for metabolite identification. We will demonstrate the dramatic improvement that a new technology, UltraPerformance LC (UPLC), brings to pharmaceutical science of metabolite ID by exploring the metabolites of acetaminophen in human urine.
A side-by-side comparison of two instrumental configurations operating under similar conditions will be made for the investigation of acetaminophen metabolites in human urine. The first system will use HPLC-MS for the analysis, and the second will employ UPLC-MS. A representative result from each of these systems is shown.
LC system: |
Waters ACQUITY UPLC System (UPLC) Waters Alliance HT System (HPLC) |
Column: |
2.1 x 100 mm, 1.7 μm ACQUITY UPLC BEH C18 Column (UPLC) 2.1 x 100 mm, 3.5 μm Symmetry C18 Column (HPLC) |
Mobile phase: |
A: 0.1% formic acid in water, B: 0.1% formic acid in acetonitrile |
Flow rate: |
500 μL/min. |
Gradient: |
0–40% B in 10 min. (UPLC) 0–20% B in 30 min. (HPLC) |
Injection volume: |
5 μL |
Sample temp.: |
10 °C |
Column temp.: |
40 °C |
MS system: |
Waters Micromass LCT Premier Mass Spectrometer |
Ionization mode: |
ESI+ |
Acquisition range: |
80–800 m/z with Woptics resolution |
Cone voltage: |
45 V |
Capillary voltage: |
3000 V |
Scan time: |
0.3 s |
Interscan delay: |
0.05 s |
LockSpray: |
25 fmol/μL Leucine enkephalin at 30 μL/min. |
As demonstrated in this simple exercise, UPLC-MS analysis offer sseveral advantages over conventional HPLC-MS for metabolite identification, namely: