• 응용 자료

Methoxyverapamil - pH 7.0, LC-MS

Methoxyverapamil - pH 7.0, LC-MS

  • Waters Corporation
Verapamil molecular structure

This is an Application Brief and does not contain a detailed Experimental section.

Abstract

This application brief highlights the analysis of methoxyverapamil using XTerra MS C18 columns.

Introduction

Methoxyverapamil has been analyzed using LC-MS in this study.

Experimental

HLC Conditions

Column:

XTerra MS C18 2.1 x 30 mm, 3.5 μm (p/n: 186000398)

Mobile phase A:

20mM NH4HCO3 in H2O, pH 7.0

Mobile phase B:

ACN

Isocratic mobile phase composition:

57%A; 43%B

Flow rate:

0.2 mL/min to MS

Injection volume:

20 μL of 100 pg/μL

Temperature:

Ambient

Detection:

MS ESI+, SIR 485.4

Instrument:

Alliance 2795 HT, Micromass ZQ

MS Conditions

MS sytem:

Micromass ZQ

Ion source:

ESI+

Capillary:

3.0 kV

Cone:

45 V

Extractor:

3.0

Source temp.:

150 °C

Desolvation temp.:

350 °C

Cone gas flow:

60 L/hr

Desolvation gas flow:

500 L/hr

LM Resolution:

15

HM Resolution:

15

Ion energy:

1

Multiplier (V):

650

Results and Discussion

Featured Products

WA20738.068, June 2002

맨 위로 돌아가기 맨 위로 돌아가기