This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates compares the new AutoTune wizard to the manual tuning capabilities of an experienced analyst, and shows that the results generated by both methods are comparable.
Creating multiple reaction monitoring (MRM) methods is an essential, yet labor-intensive part of developing sensitive and robust quantitative methodology on a tandem quadrupole LC-MS/MS system. Waters MassLynx v4.1 Software now features an integrated AutoTune Wizard with MS method development capabilities to make the calibration of the mass spectrometer, creation of appropriate tune files, and the creation of single ion recording (SIR) and MRM methods easier than ever for the analyst.
In this application brief, we compare results generated from utilizing the AutoTune wizard versus a manual tune performed by an experienced analyst for the creation of MRM methods involving ten compounds with a wide range of structural characteristics and ionization properties (Figure 1).
For both the AutoTune wizard and the manual tuning process, the LC flow was provided by an isocratic pump at 600 μL/min using a 1:1 mixture of deionized water and HPLC-grade methanol as eluent. The tuning solutions of each compound were prepared at concentrations of approximately 200 ng/mL in methanol, then infused into the LC flow at 10 μL/min. MS/MS detection was facilitated by the use of a Quattro Premier XE Tandem Quadrupole Mass Spectrometer.
The MRM Developer was run as part of the AutoTune wizard to optimize the source parameters for the compound of interest, and to find the most intense product ion and optimum collision energy. It then automatically produced an MRM method and an AutoTune report to show the optimization data used for the method generation.
In Table 1 below, a comparison of the results generated by the Auto Tune Wizard and those produced by an experienced analyst are directly compared for all ten compounds under test. A good correlation between both result sets was observed.
720001794, June 2006