        This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates analysis of Sulfonamides using Sunfire C18 and C8 Columns. Sulfonamides are class of antibiotics used to treat a wide range of bacterial infections.
The compounds analysed in this study using Sunfire C18 Column are -
 
            
        The compounds analysed in this study using Sunfire C8 150 mm Column are -
 
            
        The compounds analysed in this study using Sunfire C8 100 mm Column are -
 
            
         
            
        | Column: | SunFire C18, 4.6 x 150 mm, 5 μm | |
| Part number: | 186002559 | |
| Mobile phase A: | Water | |
| Mobile phase B: | Acetonitrile | |
| Mobile phase C: | 100 mM CH3COO-NH4+, pH 5.0 | |
| Flow rate: | 1 mL/min | |
| Injection volume: | 10 μL | |
| Sample concentration: | 10 μg/mL in water | |
| Temperature: | 30 °C | |
| Detection: | UV @ 254 nm | |
| Instrument: | Alliance 2695 with 2996 PDA | 
 
            
        | Column: | SunFire C8, 4.6 x 150 mm, 5 μm | |
| Part number: | 186002737 | |
| Mobile phase A: | Water | |
| Mobile phase B: | Acetonitrile | |
| Mobile phase C: | 100 mM CH3COO-NH4+, pH 5.0 | |
| Flow rate: | 1 mL/min | |
| Injection volume: | 10 μL | |
| Sample concentration: | 10 μg/mL in water | |
| Temperature: | 30 °C | |
| Detection: | UV @ 254 nm | |
| Instrument: | Alliance 2695 with 2996 PDA | 
Note: Method is same as for the C18.
 
            
        | Column: | SunFire C8, 4.6 x 100 mm, 3.5 μm | |
| Part number: | 186002731 | |
| Mobile phase A: | Water | |
| Mobile phase B: | Acetonitrile | |
| Mobile phase C: | 100 mM CH3COO-NH4+, pH 5.0 | |
| Flow rate: | 1 mL/min | |
| Injection volume: | 10 μL | |
| Sample concentration: | 10 μg/mL in water | |
| Temperature: | 30 °C | |
| Detection: | UV @ 254 nm | |
| Instrument: | Alliance 2695 with 2996 PDA | 
 
            
         
            
         
            
         
            
        WA41902, May 2005