This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights the analysis of verapamil by LC-MS using XTerra MS C18 columns.
Verapamil has been analyzed in this application brief.
|
Column: |
XTerra MS C18 2.1 x 30 mm, 3.5 μm (p/n: 186000398) |
|
Mobile phase A: |
0.1% NH4OH in H2O, pH 9.5 |
|
Mobile phase B: |
0.1% NH4OH in ACN, pH 9.5 |
|
Flow rate: |
0.2 mL/min to MS |
|
Isocratic mobile phase composition: |
55% A; 45% B |
|
Injection volume: |
20 μL of 100 pg/μL |
|
Temperature: |
Ambient |
|
Detection: |
MS ESI+, SIR 455.45 |
|
Instrument: |
Alliance 2795 HT, Micromass ZQ |
|
MS system: |
Micromass ZQ |
|
Source: |
ESI+ |
|
Capillary (KV): |
3.0 |
|
Cone (V): |
35 |
|
Extractor: |
3.0 |
|
RF Lens: |
0.5 |
|
Source temp.: |
150 |
|
Desolvation temp.: |
350 |
|
Cone gas flow (L/Hr): |
60 |
|
Desolvation gas flow (L/Hr): |
500 |
|
LM resolution: |
15 |
|
HM resolution: |
15 |
|
Ion energy: |
1.0 |
|
Multiplier (V): |
650 |
WA20738.117, June 2002