This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates multi-residue analysis of pesticides in avocado using AOAC QuEChERS method by UPLC-MS/MS.
| 
                     LC Conditions  | 
            |
|---|---|
| 
                     LC System:  | 
                     Waters ACQUITY UPLC System  | 
            
| 
                     Column:  | 
                     ACQUITY UPLC BEH C18, 2.1 x 100 mm, 1.7 μm  | 
            
| 
                     Column Temp:  | 
                     40 °C  | 
            
| 
                     Sample Temp:  | 
                     4 °C  | 
            
| 
                     Flow Rate:  | 
                     0.3 mL/min.  | 
            
| 
                     Mobile Phase A:  | 
                     Water + 0.1% formic acid  | 
            
| 
                     Mobile Phase B:  | 
                     Methanol + 0.1% formic acid  | 
            
| 
                     Injection Volume:  | 
                     15 μL, Partial loop injection  | 
            
| 
                     Time  | 
                     Flow Rate  | 
                     A%  | 
                     B%  | 
            
|---|---|---|---|
| 
                     0.00  | 
                     0.3  | 
                     75  | 
                     25  | 
            
| 
                     0.25  | 
                     0.3  | 
                     75  | 
                     25  | 
            
| 
                     7.75  | 
                     0.3  | 
                     5  | 
                     100  | 
            
| 
                     8.50  | 
                     0.3  | 
                     0  | 
                     100  | 
            
| 
                     8.51  | 
                     0.5  | 
                     75  | 
                     25  | 
            
| 
                     10.50  | 
                     0.5  | 
                     75  | 
                     25  | 
            
| 
                     11.0  | 
                     0.3  | 
                     75  | 
                     25  | 
            
| 
                     Instrument:  | 
                     Waters ACQUITY TQ Detector  | 
            
| 
                     Ionization:  | 
                     Positive electrospray (ESI+)  | 
            
| 
                     Acquisition:  | 
                     Multiple reaction monitoring (MRM)  | 
            
            
        720003373, March 2010