This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights the analysis of pharmaceutical residues in environmental samples using optimized Oasis MCX SPE method.
Compounds used in this application brief are:
1. Acetaminophen
2. Phenylpropanolamine
3. Salicylic acid
4. Diphenhydramine
5. Clofibric acid
6. Ethynylestradiol
7. Tamoxifen
8. Ibuprofen
|
Column: |
XTerra MS C18 4.6 x 100 mm, 3.5 μm (p/n: 186000436) |
|
Mobile phase A: |
15 mM NH4COOH, pH 4.0 |
|
Mobile phase B: |
MeOH |
|
Flow rate: |
1.0 mL/min |
|
Injection volume: |
40 μL |
|
Detection: |
MS ESI+ |
|
Instrument: |
Alliance 2695, Micromass ZQ |
|
Compound |
MW |
ESI+ |
ESI- |
|---|---|---|---|
|
Acetaminophen |
151.2 |
152(40) 110(40) |
150(20) |
|
Phenylpropanolamine |
151.2 |
152(40) 134(40) |
|
|
Salicylic acid |
138.1 |
137(17) |
|
|
Diphenhydramine |
255.3 |
256(25) 167(40) |
|
|
Clofibric acid |
214.7 |
213(17) 127(17) |
|
|
Ethynylestradiol |
296.4 |
295(40) 145(80) |
|
|
Tamoxifen |
371.5 |
372(40) |
|
|
Ibuprofen |
206.3 |
205(17) 159(30) |
Conditions for Oasis MCX 6 cc/150 mg (60 μm) Part Number 186000255
Oasis HLB Plus Part Number 186000132
WA31764.118, June 2003