This is an Application Brief and does not contain a detailed Experimental section.
This application brief highlights the analysis of chlorzoxazone in rat plasma using Oasis MAX extraction method.
Chlorzoxazone is analyzed in this application brief.
Column: |
XTerra MS C18 4.6 x 30 mm, 3.5 μm (p/n: 186000398) |
Mobile phase A: |
0.2% HCOOH |
Mobile phase B: |
ACN |
Isocratic mobile phase composition: |
45% A; 55% B |
Flow rate: |
0.2 mL/min |
Injection volume: |
20 μL |
Temperature: |
60 °C |
Detection: |
MS ESI- |
Instrument: |
Alliance 2790, Micromass Quattro Ultima |
Ion source: |
ESI- |
Source temp.: |
150 °C |
Gas cell: |
1.5e-3 mbar, 20 eV |
Desolvation temp.: |
350 °C |
Cone gas flow: |
150 L/hr |
Drying gas flow: |
600 L/hr |
Cone voltage: |
30 V |
WA31764.53, June 2003