This application brief demonstrates analysis of risperidone by LC-UV.
The compound analyzed in this study is risperidone.
|
Column: |
Xterra RP18 4.6 x 150 mm, 5 μm |
|
Part number: |
186000492 |
|
Mobile phase: |
pH 3.0: H2O/ACN/100 mM NH4COOH, pH 3.0 50:40:10 pH 7.0: H2O/ACN/100 mM NH4HCO3, pH 7.0 20:70:10 pH 10.0 H2O/ACN/100 mM NH4HCO3, pH 10.0 50:40:10 |
|
Flow rate: |
1.0 mL/min |
|
Injection volume: |
5 μL of 250 μg/mL |
|
Temperature: |
30 °C |
|
Detection: |
UV @ 280 nm |
|
Instrument: |
Alliance 2695, 2996 PDA |
|
Mobile Phase pH |
USP Tailing |
|---|---|
|
3.0 |
1.28 |
|
7.0 |
1.2 |
|
10.0 |
1.11 |
WA20738.097, June 2002