This is an Application Brief and does not contain a detailed Experimental section.
This application brief describes UPLC-MS analysis of food sugars/saccharides in beer.
Compounds analysed in beer are:
Column: |
ACQUITY UPLC BEH Amide 2.1 x 100 mm, 1.7 μm |
Part Number: |
186004801 |
Mobile Phase A: |
80/20 MeCN/H2O with 0.1% ammonium hydroxide [NH4OH] |
Mobile Phase B: |
30/70 MeCN/H2O with 0.1% ammonium hydroxide [NH4OH] |
Flow Rate: |
0.13 mL/min |
Gradient: |
10 minute gradient, 75%-45% MeCN (w/0.1% NH4OH) with 25 minute re-equilibration |
Injection Volume: |
2.0 μL (PLNO) |
Sample Concentration: |
Standards at 10 μg/mL, Beer at 50% dilution |
Sample Diluent: |
50/50 MeCN/H2O |
Column Temperature: |
35 °C |
Strong Needle Wash: |
20/80 MeCN/H2O (800 μL) |
Weak Needle Wash: |
75/2MeCN/H2O 5 (500 μL) |
Seal Wash: |
50/50 MeCN/H2O |
Instrument: |
Waters ACQUITY UPLC with SQ |
Time(min) |
Profile |
|
---|---|---|
%A |
%B |
|
0.00 |
90.00 |
10.00 |
10.00 |
30.00 |
70.00 |
10.01 |
90.00 |
10.00 |
35.00 |
90.00 |
10.00 |
Ionization Mode: |
ES- |
Capillary: |
2.8 kV |
Cone Voltage: |
25 V (fructose, glucose, maltotriose); 40V (sucrose and maltose) |
Source Temp: |
120 °C |
Desolvation Temp: |
350 °C |
Desolvation Gas Flow: |
500 L/Hr |
Cone: |
50 L/Hr |
SIR (m/z): |
179.0 (fructose, glucose); 341.1 (sucrose, maltose); 503.2 (maltotriose) |
Dwell Time: |
0.04 s |
WA60126, October 2009