This application note will discuss the use of focused gradients to maintain selectivity and resolution and to allow UPLC screening to be applied to preparative samples. This will offer the substantial time savings associated with UPLC to customers in the preparative environment.
Purification laboratories face many of the same challenges that their counterparts in analytical laboratories face: the need to increase throughput and efficiency without sacrificing quality and quantity. Successful performance of a purification lab is measured in the ability to produce pure fractions in sufficient quantities in a timely manner.
UltraPerformance LC (UPLC) has been widely accepted by chromatographers because of the improvements over HPLC in sensitivity, resolution, and speed of separations. Now scientists are beginning to explore the use of this technology in the sample screening process as a screening tool to evaluate samples prior to purification.
A typical run time for analytical screening in a preparative lab is 10 minutes. By capitalizing on the efficiency of UPLC, a 10-minute run time can be shortened to as little as one minute. This offers substantial time savings enabling for greater capacity, but also fits into the “fail fast and fail cheap” motto adopted by many pharmaceutical companies.
A standard solution of pharmaceutical-like compounds was prepared to simulate the conditions under which many purification systems operate.
LC system: |
ACQUITY UPLC System with ACQUITY UPLC Photodiode Array (PDA) Detector |
Column: |
ACQUITY UPLC BEH C18, 1.7 μm, 2.1 x 50 mm |
Injection vol.: |
2.0 μL |
Flow rate: |
0.8 mL/min, 2.1 x 50 mm |
Mobile phase A: |
0.05% Formic acid in acetonitrile |
Mobile phase B: |
0.05% Formic acid in water |
Gradient: |
Generic 5% to 95% over two minutes Focused Gradient |
LC system: |
AutoPurification System |
Column: |
XBridge Prep OBD C18,5 μm, 19 x 50 mm XBridge C18, 5 μm, 4.6 x 50 mm |
Injection vol.: |
200 μL |
Mobile phase A: |
0.05% Formic acid in acetonitrile |
Mobile phase B: |
0.05% Formic acid in water |
Flow rate: |
22 mL/min |
Gradient: |
0 to 0.25 min, 2% B to initial % B 0.25 to 1.61 min, initial % B to end % B 1.61 to 1.86 min, end % B to 95% B 1.86 to 2.71 min, 95% B 2.71 to 2.72 min, 95% B to 2% B |
MS system: |
3100 Mass Detector |
Ionization mode: |
Positive |
Switching time: |
0.05 sec |
Capillary voltage: |
3 Kv |
Cone voltage: |
60 V |
Desolvation temp.: |
350 °C |
Desolvation gas: |
500 L/Hr |
Source temp.: |
300 °C |
Acquisition range: |
150 to 700 amu |
Acquisition rate: |
5000 amu/sec |
In order to maintain the selectivity and resolution achieved by analytical analysis, the overall cycle time of a preparative analysis must be increased almost nine-fold.1 This long cycle time is not practical for most separation scientists. Therefore, we look to focused gradients to maintain selectivity and resolution in UPLC screening.
The UPLC separation of the sample shows the compound of interest eluting at 0.48 min, and is partially resolved from the peak at 0.51 min.
The separation is first directly scaled to a 19 x 50 mm XBridge Prep OBD C18 Column. The XBridge chemistry is built on the same second-generation bridged ethyl hybrid (BEH) particle as the ACQUITY UPLC BEH chemistry, in order to maintain the selectivity and resolution of the analytical analysis. To maintain the resolution and selectivity, the overall cycle time must be increased over nine-fold.
In a preparative environment, where the compound of interest is being isolated from the other components in the sample, retaining analytical resolution is not as important as isolating and collecting the compound of interest.2
A set of focused gradients can be created based on the relationship between percent composition and retention time. The system d-well time is used to determine that relationship.3
Here, in the analytical screen the mobile phase is 2% organic solvent at 0.17 minutes and 17.5% at 0.295 minutes, and so a series of gradients can be created.
The theory behind the focused gradients is the same for HPLC and for UPLC, but the time window for the UPLC gradient is much smaller.
Based on Table 1, method C is selected to isolate the compound that eluted at 0.48 min in the UPLC analysis. Using the focused gradient, the separation and isolation of the compound was carried out in three minutes.
This same methodology can be applied to the purity screening and purification of a large sample library. The ACQUITY UPLC System’s large capacity (22 384-well plates) and the rapid analysis cycle time provide the ideal tool for high throughput library screening. Data is processed and handled using AutoPurify, part of the FractionLynx Application Manager.4
AutoPurify automatically selects the samples requiring purification and the corresponding focused preparative method.
The substantial time savings associated with analytical screening can be magnified by incorporating UPLC into the analysis of the collected fractions. The collected fractions are analyzed to determine the new sample purity, and sample lists are automatically generated for each step of the process. By incorporating fraction analysis by UPLC into the workflow, the efficiency of the lab is further increased.
720002283, June 2007