This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates analysis of tetracyclines in milk.
The compound analyzed in this study is tetracyclines in milk.
The milk sample (15 mL) is diluted with 25 mL of McIlvaine (mixed citrate/phosphate, pH 4.1) buffer with added EDTA. The diluted sample is centrifuged at 8000 x g for 10 minutes at 5° C. Any floating lipid layer is removed and the remaining supernatant is processed using Oasis HLB solid-phase extraction cartridges.
For details of the buffer preparation procedure see J.AOAC Int., 79 (2), p 405 (1996).
Column: |
Nova-Pak C8, 3.9 x 150 mm, 4 μm |
Part number: |
WAT086344 |
Mobile phase: |
ACN, Methanol, 50 mm oxalic acid in water; 13:13:74 |
Flow rate: |
0.8 mL/min |
Injection volume: |
60 μL |
Detection: |
UV @ 365 nm |
WA31764.150 , June 2003