This application note describes a rapid quantitative method (LipidQuan) used to analyze the plasma from patients diagnosed with COPD and asthma in comparison to healthy controls
Chronic obstructive pulmonary disease (COPD) represents a group of lung conditions, including emphysema and chronic bronchitis, that result in significant breathing impairment (Figure 1). COPD is a non-curable disease that affects 251 million people worldwide and resulted in 3.17 million deaths in 2015 (WHO, 2017). The main cause of COPD is exposure to tobacco smoke and primarily effects middle-aged or older adults with a smoking history. Asthma and pulmonary infection are conditions linked to COPD and are driving significant increases in socio-economic cost. Recent reports have shown, for example, that the costs associated with treatment of COPD related conditions will exceed ₤11 billion per year in the UK.1 In addition to lifestyle factors such as smoking, molecular factors are known to increase the probability of developing the condition. Previous studies have shown altered lipid profiles are associated with the emergence and propagation of COPD.2,3 Although advances in mass spectrometry (MS) have allowed for more in-depth lipidomic analysis, unambiguous identification and quantification has proven difficult as lipids exhibit a high number of isomeric and isobaric species. Here, we describe the utilization of the LipidQuan4 workflow (Figure 2) to overcome these challenges and to provide a comprehensive and quantitative assessment of the lipid species involved in COPD and asthma.
The study analyzed plasma samples from 21 individuals: healthy controls healthy controls (n=6), COPD (n=9), and asthma (n=6).
Stable isotope labelled (SIL) standards (SPLASH LIPIDOMIX, Avanti Lipids, Alabaster, AL) spiked into pooled human plasma were used to generate calibration curves for quantification. In total, nine concentration levels were constructed representing all the lipid classes contained within the SIL standard. High, Middle, and Low QC samples were also prepared at 80%, 8%, and 1.275%, respectively of the highest concentration of the calibration curve.
A simple sample preparation procedure was adopted using protein precipitation with pre-cooled isopropanol (IPA) spiked with the odd Chain LIPIDOMIX (Avanti Lipids, Alabaster, AL), (1:5, plasma:IPA). Samples were vortex mixed for one minute and placed at -20 °C for 10 minutes. Samples were vortex mixed again for one minute and placed at 4 °C for two hours to ensure complete protein precipitation. The extracted samples were centrifuged at 10,300 g for 10 minutes at 4 °C and the supernatant was transferred to glass vials for LC-MS/MS analysis. Prepared samples were analyzed in duplicate for both positive and negative ionization modes.
LC system: |
ACQUITY UPLC I-Class or H-Class (Fixed Loop (FL) or Flow Through Needle (FTN)) |
Column(s): |
ACQUITY UPLC BEH Amide 2.1 × 100 mm, 1.7 μm (p/n: 186004808) |
Column temp.: |
45 °C |
Flow rate: |
0.6 mL/min |
Mobile phase: |
95:5 Acetonitrile/water + 10 mM Ammonium acetate (A) and 50:50 Acetonitrile/water + 10 mM Ammonium acetate (B) |
Gradient: |
0.1% to 20.0% B for 2 minutes, then 20% to 80% B for 3 minutes followed by 3 minutes re-equilibration |
Run time: |
8 minutes |
Injection volume: |
2 μL |
MS systems: |
Xevo TQ-XS, TQ-S, or TQ-S micro |
Ionization mode: |
ESI (+/-) |
Capillary voltage: |
2.8 kV (+) 1.9 kV (-) |
Acquisition mode: |
MRM |
Source temp.: |
120 °C |
Desolvation temp.: |
500 °C |
Cone gas flow: |
150 L/hr |
Desolvation flow: |
1000 L/hr |
Nebulizer gas: |
7 bars |
Ion guide offset 1: |
3 V |
Ion guide offset 2: |
0.3 V |
A downloadable LipidQuan Quanpedia method file that contains the LC conditions, the MS method, and the associated TargetLynx processing method (including retention times and MRM transitions) was used. The resulting data were processed with either TargetLynx or Skyline (MacCoss Lab Software, University of Washington). Statistical analysis was performed using SIMCA P+ (Umetrics, Umeå, Sweden) and further data interrogation conducted using Metaboanalyst.5
Plasma samples from three biological states of varying phenotype (healthy control, COPD patients, and asthma patients) were analyzed. Samples were prepared using a simple protein precipitation with cooled IPA before targeted LC-MS/MS data were acquired in positive and negative ion electrospray modes using the LipidQuan platform. Samples were randomized and two technical replicates per sample were acquired. Example chromatograms representing endogenous lipids for both positive and negative ion mode are shown in Figure 3. A large number of lipids were identified from the extensive LipidQuan MRM library, a library that contains highly specific fatty acyl transitions and head group fragments for increased identification confidence. Figure 4 shows the power of the LipidQuan MRM library. In this example, phosphocholines (PC), which have two potential identifications from the same precursor mass, were distinguished by the specific fatty acyl chain fragments.
Statistical analysis of the data revealed clear separation between the various cohorts (Figure 5). Validated PLS-DA models resulted in the clustering of healthy controls, COPD, and asthma patients (Figure 5A). Additional validation, through the use of permutation tests, indicated that the applied PLS-DA models did not over fit the data (R2=0.843, Q2=0.844). The loadings plot (Figure 5B), indicated that FFAs, PCs, LPCs, SMs, and Ceramides are the main contributors of statistical separation. Additional curation of the data using ANOVA/t-test revealed the 100 most statistically significant lipids for hierarchical clustering, showing differential lipid expression trends for the three cohorts (Figure 5C).
Quantification was achieved using calibration curves of plasma spiked with known concentrations of SIL standards prior to extraction. Further details of calibration ranges and acceptance criteria can be found in the LipidQuan Method Reference Guide (www.waters.com/targetedomics). Using surrogate standards prepared and analyzed under identical conditions as the control and disease samples, the quantification of endogenous lipids within the same class was achieved. Deuterated standards were used to assess linear response, with typical R2 values ranging from 0.97–0.99 for the various lipid classes in both modes of ionization. Figure 6 shows example calibration curves for LPC, PC, and PE species.
Biological relevance was explored by merging the data from all experiments and performing pathway analysis. This analysis revealed a correlation of phenotype with altered inflammatory, oxidative, and immunity processes and suggests the involvement of differential regulation of specific lipid species in signalling, metabolic, and regulatory pathways. Lipoprotein metabolism, for example, was identified as a pathway that is highly dysregulated in the COPD phenotype (Figure 7).