For forensic toxicology use only.
This application note demonstrates the sensitivity and selectivity of the expanded Forensic Toxicology Application Solution with UNIFI using negative ionization in providing a consistent comprehensive determination of cannabinoids. It can be applied as both a screen for selected cannabinoids and a method suitable for quantifying these analytes, at levels below the current EWDTS urine screening cut-off (50 ng/mL for cannabis metabolites), using a simple five-fold dilution. The excellent linear dynamic range of this system is demonstrated by simple automatically generated calibration plots.
Expanded Forensic Toxicology Application Solution with UNIFI enabling the detection and quantitation of negative ionizing cannabinoids in urine.
Cannabis is the most widely used illicit substance in the world and long-term use can lead to dependency. Cannabinoids are one of the most commonly detected classes of illegal drugs; consequently their analysis is of key importance in both forensic and workplace testing. Δ-9 tetrahydrocannabinol (THC) and cannabinol (CBN) are psychoactive elements present in the plant Cannabis sativa.1 THC produces a number of metabolites such as 11-nor-9-hydroxy-Δ9 tetrahydrocannabinol (THC-OH), but the most significant metabolite for urine drug testing is 11-nor-9-carboxy-Δ9 tetrahydrocannabinol (cTHC), which is the major metabolite eliminated in urine, as the free acid or the ester-linked β-glucuronide.2 Cannabidiol (CBD) is a major constituent of cannabis resin but is believed to have limited psychoactive properties.
The Waters Forensic Toxicology Application Solution with UNIFI currently comprises acquisition of accurate mass MSE data on an orthogonal acceleration time-of-flight mass spectrometer operating in electrospray positive ionisation mode (ESI+), followed by comparison of the data with a comprehensive library containing more than 1000 toxicologically-relevant substances.3,4,5 A number of compounds, such as the cannabinoids, also ionise in negative electrospray mode (ESI-) and the aim of the recent work was to further extend the Forensic Toxicology Application Solution to include those compounds. The new method was used to determine the presence of cannabinoids in urine, particularly at concentrations, below the current screening cut-off,6 and to compare the values obtained using this method with a recently published fully validated UPLC-MS/MS assay.7
UPLC system: |
ACQUITY UPLC I-Class (FTN) |
Column: |
ACQUITY UPLC HSS, 100Å, 1.8 μm, C18, 2.1 mm x 150 mm (p/n 186003534) |
Vials: |
Maximum Recovery Vials (p/n 186000327C) |
Column temp.: |
50 °C |
Sample temp.: |
10 °C |
Injection vol.: |
10 μL |
Flow rate: |
0.4 mL/min |
Mobile phase A: |
Water containing 0.001% formic acid |
Mobile phase B: |
Acetonitrile containing 0.001% formic acid |
Gradient: |
Isocratic at 87% A for 0.5 min, then to 5% A at 4.5 min, hold for 1 min before switching to 87% A |
Run time: |
7.5 min |
MS system: |
Xevo G2-S QTof |
Ionization mode: |
ESISource |
temp.: |
150 °C |
Desolvation temp.: |
400 °C |
Desolvation gas: |
800 L/h |
Reference mass: |
Leucine enkephalin [M-H]- m/z = 554.2620 Acquisition range m/z 50–1000 |
Scan time: |
0.1 s |
Capillary voltage: |
1.5 KV |
Cone voltage: |
20 V |
Collision energy: |
Function 1: 6 eV |
|
Function 2: ramped 10 to 40 Ev |
Reference standards THC, CBD, CBN, THC-OH, and cTHC (1 mg/mL) were purchased from LGC Standards (Teddington, UK); cTHC-glucuronide and the deuterated (d-3) analogue of cTHC (for use as internal standard; ISTD), were obtained from the same supplier at 0.1 mg/mL.
Prior to use the individual standards were diluted to 5000 ng/mL in acetonitrile and the internal standard was diluted to 100 ng/mL in 0.001% formic acid.
Bio-Rad normal control urine and Bio-Rad Liquichek Urine Toxicology Controls Level C2 and Level S10 reference urines were obtained from Bio-Rad Laboratories (Hemel Hempstead, UK).
All other chemicals used were of the highest grade available and stored according to the supplier’s instructions.
Standards (0.1 mL) were added to 0.9 mL 0.001% formic acid in a Maximum Recovery Vial and vortex-mixed to give a concentration of 500 ng/mL.
Acetonitrile (0.1 mL) was added to 0.2 mL urine and ISTD (0.7 mL). The sample was vortex-mixed, for 5 min at 1200 rpm, and then centrifuged at 8000 g for 10 min. Supernatant was transferred to a Maximum Recovery Vial.
The cannabinoids investigated in this analysis are listed in Table 1, along with their exact neutral mass, high energy fragment ions, and UPLC retention times.
The acceptance criteria for a positive identification of each analyte was as follows: three dimensional (3D) low energy ion count intensity greater than 250, retention time to be within 0.35 min of reference, and the observed precursor mass to be within 5 ppm of expected. For additional confirmation, a minimum of one diagnostic fragment ion had to be found in the high energy function. Chromatographic separation of cannabinoid standards at 100 ng/mL (50 ng/mL for cTHC-glucuronide and 500 ng/mL for cannabinol) is shown in Figure 1.
To investigate linearity, Bio-Rad control urine was spiked with cTHC and cTHC glucuronide over a range from 0 to 500 ng/mL (250 ng/mL for cTHC-glucuronide) and prepared in duplicate, as described above. The 3D peak response for each positively identified analyte trace was generated automatically during processing and referenced to the ISTD response; calibration curves were plotted using a 1/x weighting and a linear fit was applied to both analytes. The correlation coefficient (r2 value) was ≥0.995 for both analytes; a calibration curve for cTHC is shown in Figure 2. The lowest calibrator positively identified, using the criteria detailed above, was 6.25 ng/mL for cTHC-glucuronide and 12.5 ng/mL for cTHC.
Analysis of authentic urine samples Twenty-six authentic urine samples and two commercial reference urines (Bio-Rad Liquichek Level C2 and S10) were analyzed following the sample preparation method described here. The authentic samples comprised anonymized samples that had previously been quantified using a fully validated UPLC-MS/MS assay.7 The UPLC-MS/MS results for the 26 authentic samples are shown in Table 2, along with the results from the UPLC-QTof-MSE assay, showing whether cTHC or cTHC-glucuronide were positively identified based on the criteria stated previously.
The UPLC-QTof-MSE method positively identified cTHC in 11 samples and cTHC-glucuronide in 16 samples, and overall demonstrated excellent concordance with the UPLC-MS/MS data (Table 3). This demonstrated that this method can consistently detect cTHC and cTHC-glucuronide in urine, using a simple fivefold dilution, at concentrations below the current European Workplace Drug Testing Society (EWDTS) screening cut-off of 50 ng/mL for cTHC.6
Furthermore the method detected and correctly assigned cTHC in both commercial reference urines. The semi-quantitative results obtained using this method for the analysis for the Bio-Rad Liquichek Level C2 and S10 reference urines were in accordance with the manufacturer’s stated reference values, and are shown in Table 4. The additional confirmation provided by the presence of 4 high energy fragments for cTHC in the Bio-Rad Liquichek level S10 reference urine is shown in Figure 3.
This application note demonstrates the sensitivity and selectivity of the expanded Forensic Toxicology Application Solution with UNIFI using negative ionisation in providing a consistent comprehensive determination of cannabinoids. It can be applied as both a screen for selected cannabinoids and a method suitable for quantifying these analytes, at levels below the current EWDTS urine screening cut-off (50 ng/mL for cannabis metabolites), using a simple five-fold dilution. The excellent linear dynamic range of this system is demonstrated by simple automatically generated calibration plots.
CEDAM Italia, Bresso, Italy and Bianalisi Analisi Mediche, Carate Brianza, Italy for supplying the anonymized authentic urine samples.
A full validation by the user would be necessary prior to adoption in a laboratory.
720005413, May 2015