• Application Note

Gradient Separation of Amino Acids on ACQUITY UPLC BEH HILIC

Gradient Separation of Amino Acids on ACQUITY UPLC BEH HILIC

  • Waters Corporation

This is an Application Brief and does not contain a detailed Experimental section.

Experimental

Test Conditions

Column:

ACQUITY UPLC BEH HILIC, 2.1 x 50 mm, 1.7 µm

Part Number:

186003460

Mobile Phase A:

10 mM NH4COOH, 0.2% HCOOH in 50:50 ACN: H2O

Mobile Phase B:

10 mM NH4COOH, 0.2% HCOOH in 90:10 ACN: H2O

Flow Rate:

0.529 mL/min

Injection Volume:

5.0 µL

Sample Concentration:

5 µg/mL

Sample Diluent:

73:25:2 ACN:MeOH:H2O with 0.2% HCOOH and 5 µM HCl

Temperature:

30 C

Instrument:

Waters ACQUITY UPLC with SQ Mass Detector

Gradient:

Time (min)

%A

%B

0.0

0.1

99.9

1.65

0.1

99.9

4.49

99.9

0.1

4.54

0.1

99.9

5.05

0.1

99.9

MS Conditions

Ionization Mode:

ES+

Capillary:

3.0 kV

Cone:

20 V

Extractor:

3 V

RF Lens:

0.3 V

Source Temperature:

150 °C

Desolvation Temperature:

350 °C

Cone Gas Flow:

50 L/Hr

Desolvation Gas Flow:

700 L/Hr

SIR:

136.1 m/z Homocysteine

150.2 m/z Methionine

Methionine

Methionine

Homocysteine

Homocysteine
wa60134-f2
wa60134-f2

WA60134, August 2009

Back To Top Back To Top