This is an Application Brief and does not contain a detailed Experimental section.
Column: |
ACQUITY UPLC BEH HILIC, 2.1 x 50 mm, 1.7 µm |
Part Number: |
186003460 |
Mobile Phase A: |
10 mM NH4COOH, 0.2% HCOOH in 50:50 ACN: H2O |
Mobile Phase B: |
10 mM NH4COOH, 0.2% HCOOH in 90:10 ACN: H2O |
Flow Rate: |
0.529 mL/min |
Injection Volume: |
5.0 µL |
Sample Concentration: |
5 µg/mL |
Sample Diluent: |
73:25:2 ACN:MeOH:H2O with 0.2% HCOOH and 5 µM HCl |
Temperature: |
30 C |
Instrument: |
Waters ACQUITY UPLC with SQ Mass Detector |
Time (min) |
%A |
%B |
---|---|---|
0.0 |
0.1 |
99.9 |
1.65 |
0.1 |
99.9 |
4.49 |
99.9 |
0.1 |
4.54 |
0.1 |
99.9 |
5.05 |
0.1 |
99.9 |
Ionization Mode: |
ES+ |
Capillary: |
3.0 kV |
Cone: |
20 V |
Extractor: |
3 V |
RF Lens: |
0.3 V |
Source Temperature: |
150 °C |
Desolvation Temperature: |
350 °C |
Cone Gas Flow: |
50 L/Hr |
Desolvation Gas Flow: |
700 L/Hr |
SIR: |
136.1 m/z Homocysteine 150.2 m/z Methionine |
WA60134, August 2009