This is an Application Brief and does not contain a detailed Experimental section.
This analytical method can be used to monitor Propham residues in potatoes.
Propham is the active substance used as herbicides and potato sprout inhibitor. This analytical method can be used to monitor Propham residues in potatoes.
1. Add 15 g of ground potatoes into a 50 mL centrifuge tube.
2. Add 15 mL 1% acetic acid in acetonitrile and shake.
3. Add 1.5 g anhydrous sodium acetate and 6 g anhydrous magnesium sulfate (equivalent to contents of DisQuE Pouch, AOAC method).
4. Centrifuge >1500 rcf for 1 minute.
5. Take out 7.5 mL extract and dilute to 10 mL with 2.5 mL toluene.
GC System: |
Agilent 6890 |
GC Column: |
DB-5ms, 30m x 0.25mm (i.d.), 0.25 μm film. Direct connection of column to injection-port liner |
Transfer line to MS: |
300 °C |
Source temp.: |
200 °C |
Injection volume: |
1 μL splitless |
Injection port temp.: |
180 °C |
Initial temp.: |
80 °C |
Time at initial temp.: |
1 minute Then Program at 10 °C/ min to 200 °C Then at 25 °C minute to 300, hold 5 minutes |
GC-MS System: |
Waters Quattro micro GC |
Ionization mode: |
Electron Impact (70 eV) Selected-Ion Recording (SIR) |
Time (min) |
Rate Final (mL/min) |
Flow (mL/min) |
---|---|---|
0.00 |
50 |
3 |
0.50 |
50 |
3 |
0.60 |
50 |
1 |
Initial Flow: |
1 mL/min |
720002591, April 2008