This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates the isocratic separation of clonidine on Atlantis HILIC Silica Columns.
The compounds used in this study are:
|
Column: |
Atlantis HILIC Silica, 2.1 x 50 mm, 3 µm |
|
Part Number: |
186002011 |
|
Mobile Phase A: |
200 mM NH4COOH, pH 3.0 |
|
Mobile Phase B: |
ACN |
|
Flow Rate: |
0.3 mL/min |
|
Isocratic Mobile Phase: |
5% A; 95% B |
|
Injection Volume: |
5.0 µL |
|
Sample Diluent: |
75:25 ACN:MeOH |
|
Sample Concentration: |
10 ng/mL clonidine |
|
Temperature: |
Ambient |
|
Instrument: |
Waters Alliance HT System, 2795 Separations Module with Waters ZQ |
|
Ionization Mode: |
ES+ |
|
Capillary: |
2.0 kV |
|
Cone: |
40 V |
|
Extractor: |
3 V |
|
RF Lens: |
0.3 V |
|
Source Temperature: |
150 °C |
|
Desolvation Temperature: |
350 °C |
|
Cone Gas Flow: |
50 L/Hr |
|
Desolvation Gas Flow: |
700 L/Hr |
|
SIR: |
230.2 m/z |
WA64074, August 2009