This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates analysis of oxybutynin in rat plasma.
The compound analyzed in this study is oxybutynin.
Column: |
Xterra MS C18 2.1 x 30 mm, 3.5 μm |
Part number: |
186000398 |
Mobile phase A: |
100 mM NH4COOH |
Mobile phase B: |
ACN |
Isocratic mobile phase composition: |
45% A; 55% B |
Flow rate: |
0.2 mL/min |
Injection volume: |
30 μL |
Detection: |
MS ESI+ |
Instrument: |
Alliance 2790, Quattro Ultima |
Ion source: |
Electrospray Positive |
Source temperature: |
150 °C |
Gas cell: |
1.5e-3 mbar, 25eV |
Desolvation temperature: |
350 °C |
Cone gas flow: |
150 L/hr |
Drying gas flow: |
600 L/hr |
Cone voltage: |
30 V |
WA31764.115, June 2003