For forensic toxicology use only.
This is an Application Brief and does not contain a detailed Experimental section.
This application brief demonstrates analysis of drugs of abuse - LC/UV.
The compounds used in this study are –
|
Compounds |
USP Tailing |
|---|---|
|
3,4-Methylenedioxyamphetamine (MDA) |
1.08 |
|
3,4-Methylenedioxymethamphetamine (MDMA) |
1.32 |
|
Alprazolam |
1.18 |
|
Flunitrazepam |
1.15 |
|
Desmethyidiazepam |
1.14 |
|
Diazepam |
1.13 |
|
Column: |
Xterra RP18 3.0 x 100 mm, 3.5 μm |
|
Part number: |
186000420 |
|
Mobile phase A: |
5 mM NH4HCO3, pH 9.6 |
|
Mobile phase B: |
ACN |
|
Flow rate: |
0.6 mL/min |
|
Isocratic mobile phase composition: |
70% A; 30% B |
|
Injection volume: |
10 μL |
|
Sample concentration: |
10 ng/μL |
|
Temperature: |
40 °C |
|
Detection: |
UV @ 210 nm |
|
Instrument: |
Alliance 2690, 2996 PDA |
WA20738.046, June 2002